  • The Sysmex Coagulation System-2500 (CS-2500) is an automated analyzer that evaluates coagulation tests like prothrombin time (PT) and activated partial thromboplastin time (aPTT) using optical methods.
  • The study compared the CS-2500's performance with the STA-Compact analyzer by examining precision, accuracy, and reference ranges, finding that the CS-2500 had a coefficient of variation of less than 3.5% and less than 9.9% bias in accuracy.
  • Although the CS-2500 showed good agreement with the STA-Compact, slight differences were noted, suggesting that future standardization could help harmonize reference ranges across both systems.

Article Abstract

: The Sysmex Coagulation System-2500 (CS-2500) is a fully automated coagulation analyzer that uses the optical reaction method. In this study, we aimed to evaluate performance characteristics of the CS-2500 in two coagulation tests [prothrombin time (PT) and activated partial thromboplastin time (aPTT)] at our hospital laboratory. PT and aPTT measurements were performed using the CS-2500 and STA-Compact Diagnostica Stago coagulometers (STA-Compact). Then, precision, accuracy, reference range verification, and method comparison statistics were performed. In the precision study, which was performed with normal and pathologic controls for the PT-international normalized ratio (INR) and aPTT tests, all coefficient of variation% were found less than 3.5%. In the comparison study, the Passing-Bablok regression analysis demonstrated the good agreement between each analyzer for PT-INR (y = -0.081 + 1.07x and r = 0.962) and for aPTT (y = 5.498 + 0.86x and r = 0.944). Both analyzers exhibited less than 9.9% bias in the accuracy study. The reference range verification analyses revealed that the manufacturer ranges were acceptable. The verification studies of the CS-2500 coagulation system were acceptable; however, in the comparison studies, there were small differences between STA-Compact. Overall, we propose that these differences could be eliminated in future standardization studies performed to use the same reference ranges for all systems.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MBC.0000000000000920DOI Listing

Publication Analysis

Top Keywords

cs-2500 coagulation
8
reference range
8
range verification
8
cs-2500
5
analytical performance
4
performance evaluation
4
evaluation sysmex
4
sysmex cs-2500
4
cs-2500 stago
4
stago sta
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!