Abdominal aortic aneurysm (AAA) refers to the enlargement of the lower artery of the abdominal aorta, and identification of an early detection tool is urgently needed for diagnosis. In the current study, an interdigitated electrode (IDE) sensing surface was used to identify miRNA-335-5p, which reflects the formation of AAAs. The uniformity of the silica material was observed by 3D profilometry, and the chemically modified highly conductive surface improved the detection via the I-V mode. The targeted miRNA-335-5p was detected in a dose-dependent manner and based on linear regression and 3σ analyses, the sensitivity was determined to be 1 fM with a biotinylated probe. The high specificity was shown by discriminating the target sequence from noncomplementary and single- and triple-mismatched sequences. These outputs demonstrated the high-performance detection of miRNA-335-5p with good reproducibility for determination of the severity of AAA.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7214562PMC
http://dx.doi.org/10.1186/s11671-020-03331-yDOI Listing

Publication Analysis

Top Keywords

interdigitated electrode
8
determination severity
8
abdominal aortic
8
detection
4
detection microrna-335-5p
4
microrna-335-5p interdigitated
4
electrode surface
4
surface determination
4
severity abdominal
4
aortic aneurysms
4