Purpose: Dose escalation yields higher complete response to rectal tumors, which may enable the omission of surgery. Dose escalation using 50 kVp contact x-ray brachytherapy (CXB) allow the treatment of a selective volume, resulting in low toxicity and organs-at-risk preservation. However, the use of CXB devices is limited because of its high cost and lack of treatment planning tools. Hence, the MAASTRO applicator (for HDR Ir sources) was developed and characterized by measurements and Monte Carlo simulations to be a cost-effective alternative to CXB devices.
Methods And Materials: A cylindrical applicator with lateral shielding was designed to be used with a rectoscope using its tip as treatment surface. Both the applicator and the rectoscope have a slanted edge to potentially allow easier placement against tumors. The applicator design was achieved by Monte Carlo modeling and validated experimentally with film dosimetry, using the Papillon 50 (P50) device as reference.
Results: The applicator delivers CXB doses in less than 9 min using a 20375 U source for a treatment area of approximately 20 × 20 mm at 2 mm depth. Normalized at 2 mm, the dose falloff for depths of 0 mm, 5 mm, and 10 mm are 130%, 70%, and 43% for the P50 and 140%, 67%, and 38% for the MAASTRO applicator, respectively.
Conclusions: The MAASTRO applicator was designed to use HDR Ir sources to deliver a dose distribution similar to those of CXB devices. The applicator may provide a cost-effective solution for endoluminal boosting with clinical treatment planning system integration.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.brachy.2020.03.009 | DOI Listing |
Brachytherapy
August 2024
Hasselt University, Faculty of Engineering Technology, Nuclear Technological Centre (NuTeC), Hasselt, Belgium. Electronic address:
Purpose: This study aimed to develop and validate a Monte Carlo (MC) model for the Papillon+ contact x-ray brachytherapy (CXB) device, producing 50 kilovolt (kV) X-rays, specifically focusing on its application with a 25 mm diameter rectal applicator for contact therapy.
Material And Methods: The validation process involved depth dose and transverse dose profile measurements using EBT3 gafchromic films positioned in a plastic water low energy range phantom. The half-value layer (HVL) was further measured and derived from the simulated X-ray spectra.
J Contemp Brachytherapy
April 2024
Department of Urology and Radiation Oncology, the Icahn School of Medicine at Mount Sinai New York, NY, United States.
Purpose: To develop an approach to the diagnosis and treatment of prostate cancer using one platform for fusion biopsy, followed by focal gland ablation utilizing permanent prostate brachytherapy with and without a rectal spacer.
Material And Methods: Prostate phantoms containing multiparametric magnetic resonance imaging (mpMRI) regions of interest (ROI) underwent fusion biopsy, followed by image co-registration of positive sites to a treatment planning brachytherapy program. A partial hemi-ablation and both posterior lobes using a Mick applicator and linked stranded seeds were simulated.
Acta Oncol
November 2023
Department of Oncology, Aarhus University Hospital, Denmark.
Background: Extensive local disease or narrow vagina may compromise brachytherapy (BT) in patients with cervical cancer. This is the first study to analyze long-term outcomes of using 3D printed vaginal tandem-needle templates (3DP TNT) for transvaginal insertion of needles in parallel (P) or parallel and oblique (P&O) direction to the tandem.
Material And Methods: All patients treated with BT using 3DP TNT from 2015-2020 were included.
Clin Transl Radiat Oncol
January 2023
Department of Radiation Oncology (Maastro), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center+, the Netherlands.
Background And Purpose: Dose-escalation in rectal cancer (RCa) may result in an increased complete response rate and thereby enable omission of surgery and organ preservation. In order to implement dose-escalation, it is crucial to develop a technique that allows for accurate image-guided radiotherapy. The aim of the current study was to determine the performance of a novel liquid fiducial marker (BioXmark®) in RCa patients during the radiotherapy course by assessing its positional stability on daily cone-beam CT (CBCT), technical feasibility, visibility on different imaging modalities and safety.
View Article and Find Full Text PDFBrachytherapy
December 2022
Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Centre+, Maastricht, The Netherlands. Electronic address:
Introduction: The various rectal endoluminal radiation techniques all have steep, but different, dose gradients. In rectal contact brachytherapy (CXB) doses are typically prescribed and reported to the applicator surface and not to the gross tumor volume (GTV), clinical target volume (CTV) or organs at risk (OAR), which is crucial to understand tumor response and toxicity rates. To quantify the above-described problem, we performed a dose modeling study using a fixed prescription dose at the surface of the applicator and varied tumor response scenarios.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!