Objective: To analyze the impact of vacuum-assisted venous drainage (VAVD) on arterial pump flow in a simulated pediatric cardiopulmonary bypass circuit utilizing a centrifugal pump (CP) with an external arterial filter.

Methods: The simulation circuit consisted of a Quadrox-I Pediatric oxygenator, a Rotaflow CP (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set primed with Lactated Ringer's solution and packed red blood cells. Venous line pressure, reservoir pressure, and arterial flow were measured with VAVD turned off to record baseline values. Four other conditions were tested with progressively higher vacuum pressures (-20, -40, -60, and -80 mmHg) applied to the baseline cardiotomy pressure. An arterial filter was placed into the circuit and arterial flow was measured with the purge line in both open and closed positions. These trials were repeated at set arterial flow rates of 1500, 2000, and 2500 mL/min.

Results: The use of progressively higher vacuum caused a reduction in effective arterial flow from 1490±0.00 to 590±0.00, from 2020±0.01 to 1220±0.00, and from 2490±0.0 to 1830±0.01 mL/min. Effective forward flow decreased with increased levels of VAVD.

Conclusion: The use of VAVD reduces arterial flow when a CP is used as the main arterial pump. The reduction in the forward arterial flow increases as the vacuum level increases. The loss of forward flow is further reduced when the arterial filter purge line is kept in the recommended open position.An independent flow probe is essential to monitor pump flow during cardiopulmonary bypass.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7199987PMC
http://dx.doi.org/10.21470/1678-9741-2019-0311DOI Listing

Publication Analysis

Top Keywords

arterial flow
24
flow
12
forward flow
12
cardiopulmonary bypass
12
arterial
12
arterial pump
12
impact vacuum-assisted
8
vacuum-assisted venous
8
venous drainage
8
flow simulated
8

Similar Publications

Melatonin (MEL), functioning as a circulating hormone, is important for the regulation of ferroptosis in different health scenarios and acts as a crucial antioxidant in cardiovascular diseases. However, its specific function in ferroptosis related to myocardial ischemia-reperfusion injury (MIRI) remains to be fully elucidated. In our research, we utilized a rat model of MIRI induced by coronary artery ligation, along with a cell model subjected to hypoxia/reoxygenation (H/R).

View Article and Find Full Text PDF

Patent ductus arteriosus (PDA) stenting is a vital intervention for neonates with ductal-dependent blood flow, offering an attractive alternative to surgical shunt placement. Despite its benefits, the procedure poses risks such as ductal spasm, branch pulmonary artery compromise, and pseudoaneurysm formation. This report presents two complex neonatal cases with distinct outcomes.

View Article and Find Full Text PDF

Blood flow restriction (BFR) combined with low work rate exercise can enhance muscular and cardiovascular fitness. However, whether neural mechanisms mediate these enhancements remains unknown. This study examined changes in corticospinal excitability and motor cortical inhibition following arm cycle ergometry with and without BFR.

View Article and Find Full Text PDF

Mean middle cerebral artery velocity (MCAv) and the pulsatility index (PI), at rest and in response to exercise, are important markers of cerebrovascular health status in middle-aged adults, when vascular decline assumes substantial relevance. Thus, this study aimed to describe and compare the responses of MCAv and PI to incremental exercise. Two hundred and forty-eight volunteers (50-58 years, 55% women) completed a ramp test on a cycle-ergometer.

View Article and Find Full Text PDF

This study examines heat transfer and nanofluid-enhanced blood flow behaviour in stenotic arteries under inflammatory conditions, addressing critical challenges in cardiovascular health. The blood, treated as a Newtonian fluid, is augmented with gold nanoparticles to improve thermal conductivity and support drug delivery applications. A hybrid methodology combining finite element method (FEM) for numerical modelling and artificial neural networks (ANN) for stability prediction provides a robust analytical framework.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!