Axillary vein puncture guided by ultrasound vs cephalic vein dissection in pacemaker and defibrillator implant: A multicenter randomized clinical trial.

Heart Rhythm

Postgraduate Program in Cardiology and Cardiovascular Sciences-School of Medicine, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre, Brazil; Cardiology Department, Hospital de Clínicas de Porto Alegre (HCPA), Porto Alegre, Brazil.

Published: September 2020

Background: Axillary vein puncture guided by ultrasound (US-Ax) has emerged as a valid alternative access route to pacemaker and defibrillator lead insertion.

Objective: The purpose of this study was to evaluate whether US-Ax compared to cephalic vein dissection (CV) improves success and early complications in pacemaker or defibrillator implant.

Methods: This prospective, multicenter clinical trial included 88 adult patients randomized 1:1 to US-Ax (n = 44) or CV (n = 44). All procedures were performed by operators with no previous experience in axillary approach. Primary endpoint was defined as success rate. Secondary endpoints were venous access site change, time to obtain venous access, total procedural time, and early complication rate. Analyses were performed using the intention-to-treat principle.

Results: Median age was 70.5 years (58.2-79.7), and 60.2% were male. For the primary outcome, a higher success rate was observed in the axillary group (97.7% vs 54.5%; P <.001), as well as a lower rate of venous access site change (2.3% vs 40.9%; P <.001) and shorter time to obtain venous access (5 vs 15 minutes; P <.001) and procedural time (40 vs 51 minutes; P = .010), with no difference in complication rate (2.3% vs 11.4%; P =.20). In multivariate analysis, US-Ax (P <.001), single-chamber device (P = .015), and body mass index (P = .015) were independent predictors of overall success.

Conclusion: This is the first randomized trial comparing self-learned US-Ax to CV in cardiac lead implantation. Our results indicate that the axillary approach was superior in terms of success rate, time to obtain venous access and procedural time, with similar complication rate.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2020.04.030DOI Listing

Publication Analysis

Top Keywords

pacemaker defibrillator
12
axillary vein
8
vein puncture
8
puncture guided
8
guided ultrasound
8
cephalic vein
8
vein dissection
8
clinical trial
8
success rate
8
venous access
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!