Background And Objective: Evaluate depth of field (DOF) and lateral resolution (LR) on digitally assisted vitreoretinal surgery (DAVS) with different camera aperture values and TV viewing distances to determine maximal optical performance.

Patients And Methods: Five retinal surgeons objectively evaluated DOF and LR at low (10×), medium (15×), and high (20×) magnification on a conventional microscope (CM) and DAVS. DAVS was evaluated at different camera apertures (30%, 50%, and 75%) and TV viewing distances (1.2 meters [m], 1.5 m, and 2.0 m). A comparison between CM and optimized DAVS values was carried out.

Results: On DAVS, DOF significantly increased as camera aperture size was reduced (75% to 50%, P = .022; 75% to 30%, P ≤ .001; and 50% to 30%, P = .001) and as microscope magnification was reduced (high to medium, P = .009; high to low, P ≤ .001; and medium to low, P = .002). LR significantly increased as microscope magnification was increased (low to medium, P = .010; low to high, P ≤ .001; and medium to high, P = .015) and as TV display distance decreased (2.0 m to 1.5 m, P = .019; 2.0 m to 1.2 m, P = .009; and 1.5 m to1.2 m, P = .185). When comparing optimized DAVS against CM, DOF was significantly larger on DAVS (P = .019). LR on DAVS outperformed the CM, but statistical significant was not reached (P = .185).

Conclusions: The authors' study is the first to evaluate different aperture, magnification, and TV distance settings, and when optimized, DAVS can significantly outperform a CM with respect to DOF and LR. Optimized visual performance for DAVS is obtained with a 30% camera aperture and a TV viewing distance no further than 1.5 m. Maximal microscope magnification to exploit LR during macular surgery and low-to-medium magnification to increase DOF for overall vitrectomy surgery is recommended. [Ophthalmic Surg Lasers Imaging Retina. 2020;51:S15-S21.].

Download full-text PDF

Source
http://dx.doi.org/10.3928/23258160-20200401-02DOI Listing

Publication Analysis

Top Keywords

camera aperture
12
optimized davs
12
≤ 001
12
microscope magnification
12
davs
10
visual performance
8
digitally assisted
8
assisted vitreoretinal
8
vitreoretinal surgery
8
viewing distances
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!