Purpose: The recently introduced robotic x-ray systems provide the flexibility to acquire cone-beam computed tomography (CBCT) data using customized, application-specific source-detector trajectories. We exploit this capability to mitigate the effects of x-ray scatter and noise in CBCT imaging of weight-bearing foot and cervical spine (C-spine) using scan orbits with a tilted rotation axis.

Methods: We used an advanced CBCT simulator implementing accurate models of x-ray scatter, primary attenuation, and noise to investigate the effects of the orbital tilt angle in upright foot and C-spine imaging. The system model was parameterized using a laboratory version of a three-dimensional (3D) robotic x-ray system (Multitom RAX, Siemens Healthineers). We considered a generalized tilted axis scan configuration, where the detector remained parallel to patient's long body axis during the acquisition, but the elevation of source and detector was changing. A modified Feldkamp-Davis-Kress (FDK) algorithm was developed for reconstruction in this configuration, which departs from the FDK assumption of a detector that is perpendicular to the scan plane. The simulated foot scans involved source-detector distance (SDD) of 1386 mm, orbital tilt angles ranging 10° to 40°, and 400 views at 1 mAs/view and 0.5° increment; the C-spine scans involved -25° to -45° tilt angles, SDD of 1090 mm, and 202 views at 1.3 mAs and 1° increment The imaging performance was assessed by projection-domain measurements of the scatter-to-primary ratio (SPR) and by reconstruction-domain measurements of contrast, noise and generalized contrast-to-noise ratio (gCNR, accounting for both image noise and background nonuniformity) of the metatarsals (foot imaging) and cervical vertebrae (spine imaging). The effects of scatter correction were also compared for horizontal and tilted scans using an ideal Monte Carlo (MC)-based scatter correction and a frame-by-frame mean scatter correction.

Results: The proposed modified FDK, involving projection resampling, mitigated streak artifacts caused by the misalignment between the filtering direction and the detector rows. For foot imaging (no grids), an optimized 20° tilted orbit reduced the maximum SPR from ~1.5 in a horizontal scan to <0.5. The gCNR of the second metatarsal was enhanced twofold compared to a horizontal orbit. For the C-spine (with vertical grids), imaging with a tilted orbit avoided highly attenuating x-ray paths through the lower cervical vertebrae and shoulders. A -35° tilted orbit yielded improved image quality and visualization of the lower cervical spine: the SPR of lower cervical vertebrae was reduced from ~10 (horizontal orbit) to <6 (tilted orbit), and the gCNR for C5-C7 increased by a factor of 2. Furthermore, tilted orbits showed potential benefits over horizontal orbits by enabling scatter correction with a simple frame-by-frame mean correction without substantial increase in noise-induced artifacts after the correction.

Conclusions: Tilted scan trajectories, enabled by the emerging robotic x-ray system technology, were optimized for CBCT imaging of foot and cervical spine using an advanced simulation framework. The results demonstrated the potential advantages of tilted axis orbits in mitigation of scatter artifacts and improving contrast-to-noise ratio in CBCT reconstructions.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14209DOI Listing

Publication Analysis

Top Keywords

tilted rotation
8
robotic x-ray
8
x-ray scatter
8
orbital tilt
8
scans involved
8
tilt angles
8
foot imaging
8
scatter correction
8
imaging
6
tilted
5

Similar Publications

is a Gram-positive bacterium that is responsible for severe nosocomial infections. The rise of multidrug-resistant strains, which can pose significant health threats, prompts the development of new treatment interventions, and much attention has been directed at the development of prophylactic and therapeutic vaccination strategies. Capsular polysaccharides (CPs) are key protective elements of the cell wall and have been proposed as promising candidate antigens.

View Article and Find Full Text PDF

Electromagnetic Wavefront Engineering by Switchable and Multifunctional Kirigami Metasurfaces.

Nanomaterials (Basel)

January 2025

Shanghai Engineering Research Centre of Ultra Precision Optical Manufacturing, Department of Optical Science and Engineering, School of Information Science and Technology, Fudan University, Shanghai 200433, China.

Developing switchable and multifunctional metasurfaces is essential for high-integration photonics. However, most previous studies encountered challenges such as limited degrees of freedom, simple tuning of predefined functionality, and complicated control systems. Here, we develop a general strategy to construct switchable and multifunctional metasurfaces.

View Article and Find Full Text PDF

The ability of nanofluids to improve heat transmission in thermal systems is well established. This work investigates the three-dimensional theoretical behavior of Darcy-Forchheimer nanofluids in tilted magnetohydrodynamics. In this study, the Soret effect, micro-motile organisms, thermophoresis, and heat radiation are also considered.

View Article and Find Full Text PDF

Purpose: To assess the effects of modifying head position and of static ocular counter-rolling (OCR) on abduction and adduction in saccadic eye movements using a head-mounted video-oculographic device.

Study Design: A clinical observational study.

Methods: The peak velocities and amplitude gains of visually guided 12° saccades were binocularly measured in 21 healthy volunteers with their heads in the upright vertical (0°) and horizontal (± 90°, bilateral side-lying) postures, and in 6 participants with their head positions bilaterally tilted by 30°.

View Article and Find Full Text PDF

Purpose: This study aimed to evaluate the quality of clinically indicated digital dental panoramic radiographs (DPRs) of children with mixed dentition. Despite the likely widespread use of this modality, recent research detailing errors on DPRs is scarce.

Materials And Methods: A consecutive case series was performed, including 178 DPRs from patients aged 6 to 12 years.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!