Purpose: To investigate the impact of dual readout bandwidths (dBW) in a dual echo fat/water acquisition and describe a dBW-rapid acquisition relaxation enhanced, or turbo spin echo sequence where the concept is used to improve SNR by removing dead times between refocusing pulses and avoiding redundant Chemical-shift encoded.
Methods: Cramér-Rao bounds and Monte Carlo simulations were used to investigate a two-point fat/water model where the difference in bandwidths is incorporated. In vivo images were acquired at 1.5 and 3 T with the dBW-rapid acquisition relaxation enhanced, or turbo spin echo sequence. Typical bandwidth ratios were 1:2. SNR was compared with a single bandwidth sequence under identical scan parameters at 3T.
Results: Monte Carlo simulations and Cramér-Rao analysis demonstrate that number of signal averages can be improved with dual bandwidths compared to conventional single bandwidth acquisitions. The dBW-rapid acquisition relaxation enhanced, or turbo spin echo sequence can acquire images with high readout resolutions with well-conditioned sampling. An SNR improvement of 52% was measured, in line with the theoretical gain of 54%.
Conclusions: The proposed dBW-rapid acquisition relaxation enhanced, or turbo spin echo sequence is a highly SNR-efficient two-point rapid acquisition relaxation enhanced, or turbo spin echo sequence without dead times, and can acquire images at higher resolutions than current vendor-supplied alternatives.