Introduction And Aim: To compare intraocular pressure (IOP) measurements obtained with non-contact tonometry (NCT), Icare rebound tonometry (IRT) and Goldmann applanation tonometry (GAT) in pediatric cases and to examine the effect of topical anesthesia on measurements obtained using IRT.

Materials And Methods: Pediatric cases in a healthy general condition and with ophthalmic examination findings within normal limits were included in the study. IOP measurements were taken with NCT and IRT devices in all cases. Topical anesthesia was subsequently applied, and IOP measurements were then repeated using IRT and GAT tonometer devices. Differences between measurements were subjected to statistical analysis.

Results: One hundred ten eyes of 55 patients, 27 male, with a mean age of 11.44 ± 2.31 years (7-17) were included. Mean pre-anesthesia IOP values were 16.47 ± 2.89 mmHg with NCT and 17.49 ± 2.57 mmHg with IRT. Mean IOP values after topical anesthesia were 16.91 ± 2.17 mmHg with IRT and 15.51 ± 2.41 mmHg with GAT. IOP measurement values obtained with all three devices exhibited positive correlation with central corneal thickness values. Statistically significant correlation was present in terms of IOP measurement values between all three devices. However, IOP values obtained with the three devices exhibited statistically significant differences. The application of topical anesthesia caused a statistically significant decrease in IRT measurements; however, statistically significant this small change (0.58 mmHg) in recorded IOP would not be considered clinically significant.

Conclusion: IOP measurements obtained with NCT, IRT and GAT devices in the pediatric age group correlate with one another, but differ from one another in a statistically significant manner. Application of topical anesthesia affects IRT measurements; however, these small changes would not be considered clinically significant.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10792-020-01375-8DOI Listing

Publication Analysis

Top Keywords

topical anesthesia
20
iop measurements
16
iop values
12
values three
12
three devices
12
iop
10
intraocular pressure
8
pediatric age
8
age group
8
measurements
8

Similar Publications

Purpose: To investigate changes in eye alignment before and after ICL implantation in patients with myopia having corrected distance visual acuity (CDVA) of ≥0.0 logMAR.

Subjects And Methods: The medical records of 1012 patients without eye movement limitation who underwent bilateral ICL implantation were retrospectively reviewed a at the Eye Center of Sanno Hospital in Japan.

View Article and Find Full Text PDF

Studies on pain in preterm infants have usually been confined to observations of painful procedures, and information from extremely preterm infants is limited. Using registry data from a Swedish nationwide cohort, this study explored the epidemiology of pain in very preterm infants, its causes, assessments, and treatment strategies. We included liveborn infants <32 weeks' gestational age (GA) discharged between January 2020 and June 2024.

View Article and Find Full Text PDF

Several diagnostic and therapeutic endoscopic urological procedures, such as stent placement, ureteroscopy, and bladder stone lithotripsy, can be performed in a hospital, an ambulatory surgery center, in the office with IV sedation, or in the office using only topical anesthesia. The potential benefits of performing procedures in the office setting using topical anesthesia include efficiency and cost reduction. The potential harms are failure to achieve the desired outcome and patient pain.

View Article and Find Full Text PDF

To report the cosmetic, clinical, and visual outcomes of a combined surgical approach for treating a corneal/limbal dermoid using excision and a three-layered amniotic membrane graft with fibrin glue. An 18-year-old female presented with impaired vision and ocular discomfort caused by a prominent dome-shaped limbal congenital dermoid on the inferotemporal cornea, resulting in a significant aesthetic concern. A full assessment, including refraction, best-corrected visual acuity (BCVA), corneal topography, aberrometry and anterior segment OCT (AS-OCT) was conducted to plan the surgical approach.

View Article and Find Full Text PDF

Objectives: The aim of this study was to review the most effective topical anesthetic methods for reducing pain/discomfort prior to dental local anesthetic needle puncture for dental procedures in children and adolescents.

Methods: A scoping review was conducted. Individual search strategies were developed for each of the bibliographic databases (Cochrane, Embase, LILACS, LIVIVO, Pubmed, Scopus, PsyINFO, Web of Science), and in the gray literature (Google Scholar, Open Gray), comprehensively, without restrictions on language, publication data, or level of socioeconomic development of the country in which the study was conducted.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!