Background: In the field of respiratory protection for healthcare workers (HCWs), few data are available on respiratory airflow rate when HCWs are performing their work activities. The objective of this study was to assess the performance of two wearable breathing recording devices in a simulated healthcare environment.

Methods: Breathing recording devices from two different manufactures "A" and "B" were assessed using 15 subjects while performing a series of simulated healthcare work activities (patient assessment; vitals; IV treatment; changing linen; carrying weight while walking; normal breathing while standing). The minute volume (MV, L/min), mean inhalation flow (MIF, L/min), peak inhalation flow (PIF, L/min), breathing frequency (f, breaths/min), and tidal volume (TV, L/min) measured by each device were analyzed. Bland-Altman method was applied to explore the variability of devices A and B. Duncan's multiple range test was used to investigate the differences among activity-specific inspiratory flow rates.

Results: The average MV, MIF and PIF reported by device A were 23, 54, and 82 L/min with 95% upper confidence intervals (CIs) of 25, 60 and 92 L/min; the mean differences of MV, MIF and PIF presented by the two units of device A were 0.9, 1.3, and 2.8 L/min, respectively. The average values and mean differences of MV, MIF and PIF found with device B were significantly higher than device A (P<0.05), showing a high variability. During non-speech activities, the PIF/MV and MIF/MV ratios were >3.14 and >2, while with speech, the ratios increased to >6 and >3. The f during speech (15 breaths/min) was significantly lower than non-speech activities (20-25 breaths/min). Among different simulated work activities, the PIF of "patient assessment" was the highest.

Conclusions: This study demonstrated a novel approach to characterize respiratory flow for healthcare workers using an innovative wearable flow recording device. Data from this investigation could be useful in the development of future respirator test standards.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7179815PMC

Publication Analysis

Top Keywords

breathing recording
12
recording devices
12
simulated healthcare
12
work activities
12
mif pif
12
devices simulated
8
healthcare workers
8
volume l/min
8
inhalation flow
8
device l/min
8

Similar Publications

Aim: To evaluate the impact of heart rate-guided basic resuscitation compared to Helping Babies Breathe on neonatal outcomes and resuscitation practices in the Democratic Republic of the Congo.

Methods: We conducted a pre-post clinical trial comparing heart rate-guided basic resuscitation to Helping Babies Breathe in three facilities, enrolling in-born neonates ≥28 weeks gestation. We collected observational data during a convenience sample of resuscitations and extracted clinical data from the medical record for all participants.

View Article and Find Full Text PDF

Objectives: Ischemia/reperfusion (IR)-induced ventricular arrhythmia, which mainly occurs after the opening of coronary artery occlusion, poses a clinical problem. This study aims to investigate the effectiveness of pretreatment with coenzyme Q (CoQ) in combination with mitochondrial transplantation on IR-induced ventricular arrhythmias in aged rats.

Materials And Methods: Myocardial IR induction was performed by left anterior descending coronary artery occlusion for 30 min, followed by re-opening for 24 hr.

View Article and Find Full Text PDF

Background: Although substantial evidence exists regarding the treatment of pressure ulcers, there is a lack of studies demonstrating a comprehensive nursing approach for managing pressure ulcers in the ICU, particularly among patients with invasive mechanical ventilation from developing countries like Honduras. This gap in research is significant as the risk and impact of pressure ulcers on health recovery cannot be disregarded.

Objective: This study aimed to analyze Honduran nursing care for pressure ulcers in patients with invasive mechanical ventilation admitted to Intensive Care Units.

View Article and Find Full Text PDF

Prolonged exposure to high environmental temperatures results in an accumulated heat load that induces a heat stress (HS) response in dairy cattle. Heat stress compromises dairy farm profitability by reducing milk yield, altering milk composition, and hindering reproductive performance. The ability to alternate between carbohydrate and lipid sources for energy production is termed metabolic flexibility (Met Flex).

View Article and Find Full Text PDF

Background: Early and continuous exposure to painful stimuli in premature infants leads to short-and long-term complications. Listening to white noise is an accessible and inexpensive non-invasive method that can be used as a safe nursing intervention in hospitals. This study aimed to assess white noise's effect on premature Infants' physiological parameters during peripheral intravenous catheter insertion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!