Purpose: To develop and validate a Monte Carlo model of the Varian TrueBeam to study electron collimation using the existing photon multi-leaf collimators (pMLC), instead of conventional electron applicators and apertures.

Materials And Methods: A complete Monte Carlo model of the Varian TrueBeam was developed using Tool for particle simulation (TOPAS) (version 3.1.p3). Vendor-supplied information was used to model the treatment head components and the source parameters. A phase space plane was setup above the collimating jaws and captured particles were reused until a statistical uncertainty of 1% was achieved in the central axis. Electron energies 6, 9, 12, 16, and 20 MeV with a jaw-defined field of 20 × 20 cm at iso-center, pMLC-defined fields of 6.8 × 6.8 cm and 11.4 × 11.4 cm at 80 cm source-to-surface distance (SSD) and an applicator-defined field of 10 × 10 cm at iso-center were evaluated. All the measurements except the applicator-defined fields were measured using an ionization chamber in a water tank using 80 cm SSD. The dose difference, distance-to-agreement and gamma index were used to evaluate the agreement between the Monte Carlo calculations and measurements. Contributions of electron scattering off pMLC leaves and inter-leaf leakage on dose profiles were evaluated and compared with Monte Carlo calculations. Electron transport through a heterogeneous phantom was simulated and the resulting dose distributions were compared with film measurements. The validated Monte Carlo model was used to simulate several clinically motivated cases to demonstrate the benefit of pMLC-based electron delivery compared to applicator-based electron delivery.

Results: Calculated and measured percentage depth-dose (PDD) curves agree within 2% after normalization. The agreement between normalized percentage depth dose curves were evaluated using one-dimensional gamma analysis with a local tolerance of 2%/1 mm and the %points passing gamma criteria was 100% for all energies. For jaw-defined fields, calculated profiles agree with measurements with pass rates of >97% for 2%/2 mm gamma criteria. Calculated FWHM and penumbra width agree with measurements within 0.4 cm. For fields with tertiary collimation using an pMLC or applicator, the average gamma pass rate of compared profiles was 98% with 2%/2 mm gamma criteria. The profiles measured to evaluate the pMLC leaf scattering agreed with Monte Carlo calculations with an average gamma pass rate of 96.5% with 3%/2 mm gamma criteria. Measured dose profiles below the heterogenous phantom agreed well with calculated profiles and matched within 2.5% for most points. The calculated clinically applicable cases using TOPAS MC and Eclipse TPS for single enface electron beam, electron-photon mixed beam and a matched electron-electron beam exhibited a reasonable agreement in PDDs, profiles and dose volume histograms.

Conclusion: We present a validation of a Monte Carlo model of Varian TrueBeam for pMLC-based electron delivery. Monte Carlo calculations agreed with measurements satisfying gamma criterion of 1%/1 mm for depth dose curves and 2%/1 mm for dose profiles. The simulation of clinically applicable cases demonstrated the clinical utility of pMLC-based electrons and the use of MC simulations for development of advanced radiation therapy techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14194DOI Listing

Publication Analysis

Top Keywords

monte carlo
36
carlo model
20
carlo calculations
16
gamma criteria
16
electron delivery
12
model varian
12
varian truebeam
12
dose profiles
12
electron
10
carlo
9

Similar Publications

Theory and simulations are used to demonstrate implementation of a variational Bayes algorithm called "active inference" in interacting arrays of nanomagnetic elements. The algorithm requires stochastic elements, and a simplified model based on a magnetic artificial spin ice geometry is used to illustrate how nanomagnets can generate the required random dynamics. Examples of tracking and PID control are demonstrated and shown to be consistent with the original stochastic differential equation formulation of active inference.

View Article and Find Full Text PDF

Multi-item retro-cueing effects refer to better working memory performance for multiple items when they are cued after their offset compared to a neutral condition in which all items are cued. However, several studies have reported boundary conditions, and findings have also sometimes failed to replicate. We hypothesized that a strategy to focus on only one of the cued items could possibly yield these inconsistent patterns.

View Article and Find Full Text PDF

Towards Rational Design of Confined Catalysis in Carbon Nanotube by Machine Learning and Grand Canonical Monte Carlo Simulations.

Angew Chem Int Ed Engl

December 2024

State Key Laboratory of Catalysis, Dalian National Laboratory for Clean Energy, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, 116023, P. R. China.

The microenvironment is recognized to be as crucial as active sites in heterogeneous catalysis. It was found that the catalytic activity of a set of chemical reactions can be significantly influenced by the confined space of carbon nanotubes (CNTs), with some reactions showing superior activity, while others experience a negative impact. The rational design of confined catalysis must rely on the accurate insights of confined microenvironment.

View Article and Find Full Text PDF

Volume electron microscopy (vEM) enables biologists to visualize nanoscale 3D ultrastructure of entire eukaryotic cells and tissues prepared by heavy atom staining and plastic embedding. The highest resolution vEM technique is focused ion-beam scanning electron microscopy (FIB-SEM), which provides nearly isotropic (~5-10 nm) spatial resolution at fluences of > 10,000 e /nm . However, it is not clear how such high resolution is achievable because serial block-face (SBF) SEM, which incorporates an in-situ ultramicrotome instead of a Ga FIB beam, results in radiation-induced collapse of similar specimen blocks at fluences of only ~20 e /nm .

View Article and Find Full Text PDF

A significant advancement in synthetic biology is the development of synthetic gene circuits with predictive Boolean logic. However, there is no universally accepted or applied statistical test to analyze the performance of these circuits. Many basic statistical tests fail to capture the predicted logic (OR, AND, etc.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!