Purpose: Interventional procedures involving radionuclides (e.g., radioembolization) would benefit from single-photon emission computed tomography (SPECT) performed in the intervention room because the activity distribution could be immediately visualized. We believe it might be possible to perform SPECT with the C-arm cone beam computed tomography (CBCT) scanner present in the intervention room by equipping the x-ray flat panel detector with a collimator. The purpose of this study is to demonstrate the approach and to investigate the achievable SPECT reconstruction quality.

Methods: A proof-of-concept experiment was performed to evaluate the possibility of nuclear imaging with an x-ray flat panel detector. The experiment was digitally replicated to study the accuracy of the simulations. Three flat panel configurations (with standard hardware and reconstruction methodology, with sophisticated reconstruction methodology, and with expected future hardware) and a conventional gamma camera were evaluated. The Jaszczak and the NEMA IQ phantom (filled with Tc) were simulated and assessed on resolution and contrast-to-noise ratio (CNR).

Results: The proof-of-concept experiment demonstrated that nuclear images could be obtained from the flat panel detector. The simulation of the same configuration demonstrated that simulations could accurately predict the flat panel detector response. The CNR of the 37 mm sphere in the NEMA IQ phantom was 22.8 ± 1.2 for the gamma camera reconstructions, while it was 11.3 ± 0.7 for the standard flat panel detector. With sophisticated reconstruction methodology, the CNR improved to 13.5 ± 1.4. The CNR can be expected to advance to 18.1 ± 1.3 for future flat panel detectors.

Conclusions: The x-ray flat panel detector of a CBCT scanner might be used to perform nuclear imaging. The SPECT reconstruction quality will be lower than that achieved by a conventional gamma camera. The flat panel detector approach could, however, be useful in providing a cost-effective alternative to the purchase of a mobile SPECT scanner for enabling interventional scanning.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7496965PMC
http://dx.doi.org/10.1002/mp.14191DOI Listing

Publication Analysis

Top Keywords

flat panel
40
panel detector
32
x-ray flat
16
nuclear imaging
12
reconstruction methodology
12
gamma camera
12
flat
10
panel
10
imaging x-ray
8
detector
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!