Establishment of national diagnostic reference levels for radiotherapy computed tomography simulation procedures in Slovenia.

Eur J Radiol

University of Ljubljana, Faculty of Health Sciences, Medical Imaging and Radiotherapy Department, Zdravstvena pot 5, 1000, Ljubljana, Slovenia. Electronic address:

Published: June 2020

Purpose: To propose national diagnostic reference levels (DRLs) for radiotherapy (RT) computed tomography (CT) localization purposes, compare both CT units used in the largest RT department in the country and to compare gathered results with other published DRLs in order to discover any need of optimization.

Methods: In total, 1631 patient data (time spend of 4 months) regarding sex, examination type, total dose-length product (DLP) and CTDI was collated on two CT units. Those simulation procedures account for more than 80 % of all simulation procedures performed nationwide. Then, total DLP and CTDI was calculated and mean, median and 3 quartile for both units together were presented to determine national DRLs for simulation procedures. The same data was later compared between both units to discover any potential need for optimization.

Results: 3 quartile values of DLP for abdomen, breast, chest, head, head and neck, pelvis and spine were 1116.2, 606.6, 832.4, 1942.4, 969.2, 677.1 and 1042.4 mGy∙cm, respectively. 3 quartile CTDI values for the same sequence of procedures were 18.7, 13.3, 19.2, 76.9, 22.6, 17.9 and 22.2 mGy, respectively. Among the two units, the mentioned dose values were on average significantly higher on one CT unit than on the other unit.

Conclusions: When comparing collected dose values with other studies, RT CT DRLs showed that radiation doses from our institution were similar or even lower. Some variations were found between both CT units in certain protocols, so exposure parameters should be reviewed and optimized.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2020.108979DOI Listing

Publication Analysis

Top Keywords

simulation procedures
16
national diagnostic
8
diagnostic reference
8
reference levels
8
radiotherapy computed
8
computed tomography
8
dlp ctdi
8
dose values
8
units
6
procedures
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!