Biotin taken orally can interfere with some diagnostic immunoassays, including those for thyroid hormones, ferritin, and markers of infectious disease. Assays affected are ones that use streptavidin-biotin in their design. The goal of our study was to examine the effect of biotin concentrations of up to 1200 ng/mL on three serological assays performed on VITROS 3600 system, Immunoglobulin M (IgM) antibodies to hepatitis A virus (anti-HAV), total anti-HAV, and IgM antibodies to hepatitis B virus core antigen (anti-HBc), by spiking serum samples with variable amounts of biotin. No false-negative results were generated with either concentration of biotin for total anti-HAV (65/65). Likewise, biotin caused no false-positive IgM anti-HAV results (59/59) with either concentration of biotin; however, 6.7% false negativity was found for IgM anti-HAV when samples were spiked with 1200 ng/mL of biotin. Conversely, 100% false positivity (30/30) was produced by biotin interference in total anti-HAV negative specimens with both concentrations of biotin. False negativity rate was 87.5% in IgM anti-HBc positive samples when biotin levels were at 1200 ng/mL. These data show that individuals taking biotin-containing supplements may test false-positive in some serologic assays using streptavidin-biotin chemistries. Further studies are warranted to determine the extent of biotin interference resulting in false-positive and negative results and their impact, if any, on surveillance and diagnostic settings.