Background: Subxyphoid active left ventricular epicardial (LVE) lead implants or VT ablation are attractive but remain a challenge due to concerns of coronary artery damage. We aimed to see if Doppler-guided positioning could permit safe LVE lead placement without coronary angiography. We evaluated the feasibility of a Doppler flow-guided subxyphoid epicardial screw-in lead fixation in a swine model.

Methods: Acute subxyphoid access to the pericardial space was performed in an anesthetized swine model using a deflectable sheath and a modified needle-derived Doppler flow meter. The audio signal and visual display from the Doppler flow meter were recorded. Coronary angiography was performed to verify the catheter location. A SelectSecure Model 3830 lead (Medtronic) was used to assess pacing in the procedure.

Results: In both of two swine, the deflectable catheter was inserted into pericardial space via subxyphoid access. The tip of the deflectable catheter with the Doppler was directed to several locations, from quiet (no nearby coronary artery expected) to typical rhythmic pulsatile sound locations which were maximal when superimposed on a coronary artery. Repeated coronary angiograms confirmed the expected findings. A 3830 active lead was fixed into a quiet location for LVE pacing, and confirmed by angiography as distant from a coronary artery.

Conclusions: Doppler-guided subxyphoid epicardial screw-in lead placement is feasible once the catheter tip is directed and stabilized in a desired LVE location. This obviates the need for repeated (or any) coronary angiography. The Doppler-guided subxyphoid epicardial procedure may also be applicable for epicardial ventricular arrhythmia ablation procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.14496DOI Listing

Publication Analysis

Top Keywords

coronary artery
16
doppler-guided subxyphoid
12
lead placement
12
coronary angiography
12
subxyphoid epicardial
12
coronary
9
artery damage
8
left ventricular
8
ventricular epicardial
8
lve lead
8

Similar Publications

To investigate the correlation between the density and volume of epicardial adipose tissue(EAT)and acute coronary syndrome (ACS). This study included 355 subjects (mean age: 60.65 ± 9.

View Article and Find Full Text PDF

Background: Discrepancies often exist between patient-reported symptoms and diagnostic test findings in patients with suspected obstructive coronary artery disease (CAD).

Objectives: To quantify the association between patient-reported symptoms and the diagnosis of severe stenosis in patients with suspected obstructive CAD.

Methods: Two large-scale cohorts of patients with new-onset symptoms suggestive of obstructive CAD were evaluated.

View Article and Find Full Text PDF

Beating-heart CABG in patients with LV dysfunction can provide the best of all words by limiting myocardial injury purported by cardioplegic arrest. Complete revascularization is possible and graft numbers are not different when compared to arrested heart CABG. Furthermore, beating-heart CABG more often reduces the need for intraoperative and postoperative mechanical support reducing the complications and costs associated with these devices.

View Article and Find Full Text PDF

Background: Intravascular lithotripsy (IVL) has an excellent efficacy and safety profile in the treatment of calcified coronary lesions during percutaneous coronary intervention (PCI). However, data regarding its use on left main (LM) lesions are still limited.

Objective: We aimed to analyze the technical success and 1-year clinical outcomes in calcified LM lesions treated with IVL.

View Article and Find Full Text PDF

Biomagnetic fluid dynamics (BFD) is an emerging and promising field within fluid mechanics, focusing on the dynamics of bio-fluids like blood in the presence of magnetic fields. This research is crucial in the medical arena for applications such as medication delivery, diagnostic and therapeutic procedures, prevention of excessive bleeding, and treatment of malignant tumors using magnetic particles. This study delves into the intricacies of blood flow induced by cilia, carrying trihybrid nanoparticles (gold, copper, and titania), within a catheterized arterial annulus under a robust magnetic field.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!