Purpose: This study focused on determining risks from stereotactic radiotherapy using flattening filter-free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs.
Methods: A retrospective review was performed. Subsequently, a phantom study was performed using 30 fully functional explanted CIEDs from two different manufacturers. Irradiation was carried out in a slab phantom with 6-MV and 10-MV FFF beams. First, a repetition-rate test (RRT) with a range of beam pulse frequencies was conducted. Then, multifraction SBRT (48 Gy/4 Fx) and single-fraction SBRT (28 Gy/1 Fx) treatment plans were used for lung tumors delivered to the phantom.
Results: Between 2014 and 2018, 13 cases were treated with an FFF beam (6 MV, 1400 MU/min or 10 MV, 2400 MU/min), and 15 cases were treated with a flattening filter (FF) beam (6 MV, 600 MU/min). All the devices were positioned outside the treatment field at a distance of more than 5 cm, except for one case, and no failures were reported due to SBRT/SRS. In the phantom rep-rate tests, inappropriate sensing occurred, starting at a rep-rate of 1200 MU/min. Cardiac implantable electronic device anomalies during and after delivering VMAT-SBRT with a 10-MV FFF beam were observed.
Conclusions: The study showed that caution should be paid to managing CIED patients when they undergo SBRT using FFF beams, as it is recommended by AAPM TG-203. Correspondingly, it was found that for FFF beams although there is small risk from dose-rate effects, delivering high dose of radiation with beam energy greater than 6 MV and high-dose rate to CIEDs positioned in close vicinity of the PTV may present issues.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7324704 | PMC |
http://dx.doi.org/10.1002/acm2.12873 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!