Introduction: Hinged knee megaprostheses are mainly used for reconstruction after tumor resection. They may incur complications, but this has not been assessed in the French literature, except in small series at short follow-up. We therefore conducted a large-scale nationwide multicenter retrospective study with a minimum 5 years' follow-up. The objectives were (1) to compare survival between distal femoral and proximal tibial reconstruction prostheses, and (2) to analyze complications and failure.
Hypothesis: Distal femoral hinged reconstruction prostheses show longer survival with fewer complications than proximal tibial prostheses.
Material And Methods: One hundred sixty-one patients were included: 118 in the distal femoral group, and 43 in the proximal tibial group. Tumors were mostly osteosarcomas (90 cases) or chondrosarcomas (31 cases). Mean age was 37 years (range, 12-86 years). Complications were assessed on the Henderson classification. Failure was defined by prosthesis anchor exchange or amputation.
Results: At a mean 9 years' follow-up (range, 5-23 years), implant survival was longer in the distal femoral group: 5- and 10-year survival, 84% [95% CI, 75-89] and 70% [95% CI, 59-79] versus 74% [95% CI, 69-85] and 43% [95% CI, 23-61] (p=0.02). Revision surgery for complications mainly concerned aseptic loosening (19%, 30 cases) or deep infection (16%, 25 cases) and more often involved the proximal tibia (65% vs. 43%, 28 vs. 51 cases; OR 2.4 [95% CI, 1.2-5.1]; p=0.02).
Discussion: Hinged knee reconstruction prosthesis is a solution in tumoral pathology, but with a high risk of complications (loosening and infection) and a higher failure rate in the proximal tibial reconstruction prosthesis.
Level Of Evidence: IV, case series.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.otsr.2019.11.027 | DOI Listing |
J Clin Med
December 2024
Department of Orthopaedic Surgery, Institute of Medicine, University of Tsukuba, 1-1-1, Tennodai, Tsukuba 305-8575, Ibaraki, Japan.
Proximal femoral fractures are particularly common in older adults, and cases requiring conversion to total hip arthroplasty may arise because of treatment failure or osteoarthritis. Fractures around the distal screw removal holes can be problematic. This study aimed to analyze the relationship between stem length and femoral stress distribution to determine the optimal stem length.
View Article and Find Full Text PDFSci Rep
January 2025
Robotics Institute, Carnegie Mellon University, 5000 Forbes Avenue, Pittsburgh, PA, 15213, USA.
Patient-specific templating (PST), which is a sister procedure to patient-specific instrumentation (PSI) but hospital-based, is relatively less complex and less expensive than robotics and navigation. However, there are some concerns about the PST including the process of preoperative planning, 3D printing and material, positioning of PST intraoperatively, availability, and clinical value. The purpose of this study was to validate the technical accuracy and reliability of the PST technique in the lab and to report the outcomes of clinical application.
View Article and Find Full Text PDFPurpose: Tibial rotational deformity is a known risk factor for patellofemoral joint (PFJ) disorders. However, it is commonly associated with other abnormalities which affect the PFJ. The purpose of this study was to describe the prevalence of associated factors known to affect PFJ in patients undergoing rotational tibial osteotomy and their implication for the correction level.
View Article and Find Full Text PDFAnn Thorac Surg Short Rep
December 2024
Division of Cardiac Surgery, Inova Heart and Vascular Institute, Inova Health Systems, Falls Church, Virginia.
Background: DeBakey type I aortic dissections (AD) are most frequently treated with hemiarch repair. A subset of patients demonstrates persistent distal end-organ ischemia secondary to persistent true lumen (TL) compression. We describe the use of bare metal stent grafting across the residual arch dissection with the Zenith Dissection Endovascular Stent (ZDES, Cook Medical) in 7 patients with type I AD that was repaired in a hemiarch configuration with a compromised distal TL and organ malperfusion.
View Article and Find Full Text PDFInjury
December 2024
Golden Jubilee Medical Center, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Background: Severe metaphyseal comminution and sizable bone defect of the distal femur are high risks of fixation failure. To date, no exact magnitude of comminution and bone loss is determined as an indication for augmentation of fixation construct. The present study aimed to investigate the influence of metaphyseal gap width, working length, and screw distribution on the stability of the fixation construct.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!