: Anterior chamber (AC) flare is a key sign for anterior uveitis. New instrument-based techniques for measuring AC flare can offer automation and objectivity. This review aims to identify objective instrument-based measures for AC flare.: In this systematic review, we identified studies reporting correlation between instrument-based tests versus clinician AC flare grading, and/or aqueous protein concentration, as well as test reliability.: Four index tests were identified in 11 studies: laser-flare photometry (LFP), optical coherence tomography, ocular flare analysis meter (OFAM) and the double-pass technique. The correlation between LFP and clinician grading was 0.40-0.93 and 0.87-0.94 for LFP and protein concentration. The double-pass technique showed no correlation with clinician grading and insufficient information was available for OFAM.: LFP shows moderate to strong correlation with clinician grading and aqueous protein concentration. LFP could be a superior reference test compared to clinician AC flare grading for validating new index tests.

Download full-text PDF

Source
http://dx.doi.org/10.1080/09273948.2019.1709650DOI Listing

Publication Analysis

Top Keywords

protein concentration
12
clinician grading
12
instrument-based tests
8
anterior chamber
8
chamber flare
8
systematic review
8
identified studies
8
clinician flare
8
flare grading
8
aqueous protein
8

Similar Publications