Background: Serum triglyceride concentrations are measured as total glycerides content in many Western countries. In Japan, glycerol-blanked triglycerides (TG-GB) are measured to identify postprandial hypertriglyceridaemia and to minimize the influence of glycerol formulation on serum triglyceride values. However, TG-GB measurements have not been standardized. Therefore, we developed an efficient quantification system for total glycerides and free glycerol that allows the calculation of TG-GB concentrations.

Methods: We measured total glycerides and free glycerol in human serum by using gas chromatography-isotope dilution mass spectrometry and compared its performance to the reference method of the US Centers for Disease Control and Prevention (CDC).

Results: Our practical method of total glycerides and free glycerol quantification achieved excellent precision for both within-run and among-run coefficients of variation (<1.5% and <2.7%, respectively), with an average recovery of 99.8% for free glycerol. However, we noted an average %bias of -0.26% for total glycerides and -3.15% for free glycerol between our TG-GB method and the CDC reference method.

Conclusions: This practical method of total glycerides and free glycerol quantification enables traceability assessment of TG-GB measurements. Differences between the output values of TG-GB and the CDC reference method might result from the differences in free glycerol values.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0004563220921884DOI Listing

Publication Analysis

Top Keywords

total glycerides
16
glycerides free
12
free glycerol
12
triglyceride concentrations
8
gas chromatography-isotope
8
chromatography-isotope dilution
8
dilution mass
8
mass spectrometry
8
serum triglyceride
8
measured total
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!