Purpose: To present on the commissioning of an automated brachytherapy plan checker (BPC) for the evaluation of high-dose-rate brachytherapy treatment plans in support of standardized workflows and patient safety.
Methods And Materials: A BPC was developed using an applications programming interface in a commercial treatment planning system based on different inputs (e.g., regulations, professional society recommendations, and user feedback) and leveraged our experience with an in-house developed external beam plan checker. The BPC was commissioned using a comprehensive suite of test plans with known errors and anonymized clinical plans.
Results: During commissioning, the BPC was successfully executed on a total of 87 test plans. Commissioning tests spanned a range of treatment sites and evaluated that pass and fail states were correct. Administration settings were changed in a nonclinical database to evaluate tests involving the source and afterloader. Clinical testing of the BPC was then performed in parallel with a manual review process before clinical implementation.
Conclusions: To commission the BPC for clinical use, a comprehensive suite of test plans was developed and used to ensure the BPC correctly detected and reported errors. A summary of the test plans is presented to help guide users developing similar automated tools. The BPC represents a process-improvement initiative designed to reduce errors and improve safety for brachytherapy patients. By using a comprehensive test suite for commissioning, tests are available for periodic quality assurance and after software upgrades.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.brachy.2020.02.003 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!