Background And Aim: Probe-based confocal laser endomicroscopy (pCLE) provides real-time microscopic visualisation. Our aim was to compare the diagnostic accuracy of pCLE with standard biopsies in patients with visible oesophageal or gastric lesions.
Methods: This was a single-centre, prospective, pathologist-blinded study. Patients underwent high-resolution endoscopy, and lesions were examined by pCLE followed by standard biopsies. A definitive diagnosis was determined from resection specimen. Main outcomes were overall diagnostic accuracy, sensitivity, specificity and positive and negative predictive values.
Results: We examined 74 lesions in 67 patients. Definitive diagnoses revealed 34 malignant and 40 non-malignant lesions. pCLE diagnosis was correct in 89.2% (66/74), while diagnosis based on biopsy was correct in 85% (57/67; = 0.6). The overall diagnostic accuracy of biopsies was 85% (76-94%) and that of pCLE was 89% (79-96%). pCLE correctly diagnosed malignant lesions, comprising oesophageal adenocarcinoma, oesophageal squamous-cell cancer or gastric adenocarcinoma, in 88.2% (30/34) of cases, while biopsy was correctly diagnosed in 75.9% (22/29; = 0.3). Sensitivity and specificity to diagnose a malignant lesion were 75.9% (95% confidence interval (CI) 56-89%) and 100% (95% CI 90-100%) for biopsies and 88.2% (95% CI 72-97%) and 92% (95% CI 79-98%) for pCLE. No differences between biopsies and pCLE were found with regard to sensitivity, specificity to diagnose dysplastic and benign lesions ( > 0.2).
Conclusion: pCLE provides satisfactory diagnostic accuracy comparable with standard biopsies in patients with oesophageal or gastric lesions. ClinicalTrials.gov identifier: NCT0292049).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7226690 | PMC |
http://dx.doi.org/10.1177/2050640620904865 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!