Biomechanical comparison of a 3.5 mm anterior locking plate to cannulated screws with anterior tension band wiring in comminuted patellar fractures.

Injury

Department of Orthopedics and Trauma Surgery, Medical Center - Albert-Ludwigs-University of Freiburg, Faculty of Medicine, Albert-Ludwigs-University of Freiburg, Germany; G.E.R.N. Tissue Replacement, Regeneration & Neogenesis, Department of Orthopedics and Trauma Surgery, Medical Center - Albert-Ludwigs-University of Freiburg, Faculty of Medicine, Albert-Ludwigs-University of Freiburg, Germany.

Published: June 2020

Purpose: To date, surgically treated multifragmentary patellar fractures are still associated with high rates of complications, such as i.e. secondary fracture displacement. Osteosynthesis is most frequently performed with screws and cerclages. To increase primary stability, locking plates have been introduced. However, there is still a lack of biomechanical data supporting the superiority of plate fixation compared to screw fixation with cerclages in these cases. The goal of the present study was to conduct biomechanical comparison of these two techniques under dynamic loading conditions.

Methods: A standardized 34-C3 fracture was created in eight pairs of human cadaveric knee joints. Following a randomization protocol, they were fixed with a 3.5 mm anterior locking plate (LP) or cannulated screws with anterior tension band wiring (hybrid osteosynthesis, HO).Subsequently, all constructs were tested for 100 cycles from 90° of knee-flexion to full extension by applying a pulling force to the quadriceps tendon. Outcome parameters were fracture displacement after one cycle, after 100 cycles and implant loosening. Failure was defined as fracture displacement > 2 mm.

Results: Biomechanical testing showed significantly less fracture displacement following LP compared to HO both after the first (p = 0.042) and after 100 cycles (p = 0.025). The difference in loosening was significant as well (p = 0.017). Following HO, 5/8 constructs failed during cyclic loading. There was no failure in the LP group. In the HO group, loosening correlated with bone mineral density (R = - 0.857) which was not observed in the LP group (R = - 0.429).

Conclusion: Anterior locking plate osteosynthesis of comminuted patellar fractures biomechanically provides better primary stability compared to cannulated screws with anterior tension band wiring.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2020.03.030DOI Listing

Publication Analysis

Top Keywords

anterior locking
12
locking plate
12
cannulated screws
12
screws anterior
12
anterior tension
12
tension band
12
band wiring
12
patellar fractures
12
fracture displacement
12
100 cycles
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!