Background: The Echelon circular™ powered stapler (ECP stapler) obviates the need for manual firing of conventional circular staplers during the construction of a colorectal anastomosis, but has not been evaluated clinically. The aim of this study was to perform a clinical evaluation of this stapler.

Methods: A retrospective review of the initial clinical experience of a single surgeon using the ECP stapler for left-sided colorectal anastomosis construction during elective colorectal resections for benign and malignant disease was conducted by analyzing results from a prospectively maintained study database. Additionally, four attending colorectal and/or general surgeons who had performed ≥ 5 colorectal operations with the ECP stapler were invited to complete an anonymous online survey to subjectively assess the user experience with the device. Statistical analysis was conducted using Microsoft Excel Version 15.33.

Results: Seventeen patients underwent left-sided anastomotic reconstruction using the ECP stapler. All donuts (proximal and distal) were intact. Anastomotic integrity was evaluated using the air-leak test utilizing flexible video sigmoidoscopy. No leaks were observed, although one patient (5.9%) developed a postoperative pelvic abscess. The anonymous survey was completed by all four surgeons. Subjective evaluation of the ECP stapler suggests that the overall stapling quality, overall device ease-of-use, and the overall perception of anastomotic quality as above average when compared to manual 'end-to-end anastomosis' (EEA) stapling devices.

Conclusions: In an initial clinical evaluation of the ECP stapler, the safety and ease-of-use of the device appears to be satisfactory. Powered stapling and the design of '3D stapling' may provide advantages over manual systems, and may improve the construction quality of left-sided colorectal anastomosis.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10151-020-02162-4DOI Listing

Publication Analysis

Top Keywords

ecp stapler
24
colorectal anastomosis
16
initial clinical
12
clinical experience
8
stapler
8
clinical evaluation
8
left-sided colorectal
8
evaluation ecp
8
colorectal
7
ecp
6

Similar Publications

Background: The most common postoperative complication in malignant rectal surgery is anastomotic leakage (AL). AL after anterior or low anterior resection in rectal tumors is a fatal postoperative complication. Recently, the first automated suture circular stapler, which is expected to reduce the incidence of AL, (J&J).

View Article and Find Full Text PDF

Background: Circular stapler anastomosis is a common surgical procedure. Despite technological advancements, anastomotic leak remains a postoperative concern. Assessment of new technologies is impeded by variations in test methods and analysis, precluding outcome reproducibility and direct comparisons of results across studies.

View Article and Find Full Text PDF

Aims: This study aimed to examine the economic and clinical benefits of a new powered circular stapler for left-sided colorectal construction in China.

Methods: A decision analysis model was constructed for a cohort of adult patients who underwent left-sided colorectal anastomoses, using either the Echelon Circular Powered (ECP) stapler) or the conventional manual circular staplers (MCS). The complications rates and healthcare resource utilization in the ECP cohort were obtained from the single-arm ECP trial (NCT03326895).

View Article and Find Full Text PDF

Purpose: To estimate the cost impact of using the ECHELON CIRCULAR™ Powered Stapler (ECP) compared with manual circular staplers (standard of care, SOC) among patients undergoing colectomy procedures that involve left-sided anastomosis.

Methods: A US hospital-based budget impact model was developed to estimate the impact of ECP in reducing the surgical complication of anastomotic leak. The incremental acquisition cost of ECP vs SOC was compared to the net potential savings from reduced complication costs.

View Article and Find Full Text PDF

Background: This was a retrospective, matching-adjusted indirect comparison of clinical outcomes between patients from a single-arm trial of the ECHELON CIRCULAR™ Powered Stapler (ECP) and those from a historical cohort of patients who underwent left-sided colorectal resection using conventional manual circular staplers, extracted from the Premier Healthcare Database.

Methods: Patients in the ECP trial cohort were propensity score matched to those in the historical cohort through nearest neighbor matching. Outcomes included 30-day readmission rates; length of stay (LOS) for the index admission; rates of anastomotic leak, pelvic abscess, ileus/small bowel obstruction, infection, bleeding, and stoma creation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!