The calibration of an underwater spectrometer, consisting of a CeBr crystal, is performed for sediment sample measurements, via experimental data and Monte Carlo simulations. The system was implemented for activity concentration measurements in marine sediment samples in the laboratory. The results were compared with corresponding high resolution measurements and showed that the system provides accurate results, but relatively high uncertainties. Different measurement geometries are theoretically studied via Monte Carlo simulations in order to improve the system performance for such measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2020.109124DOI Listing

Publication Analysis

Top Keywords

sediment samples
8
monte carlo
8
carlo simulations
8
measurements
5
calibration cebr
4
cebr based
4
based γ-spectrometer
4
γ-spectrometer onsite
4
onsite laboratory
4
laboratory radioactivity
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!