Small-Incision Lenticule Extraction (SMILE) for the Correction of Myopia with Astigmatism: Outcomes of the United States Food and Drug Administration Premarket Approval Clinical Trial.

Ophthalmology

Department of Ophthalmology, University of California, San Francisco, San Francisco, California; Carl Zeiss Meditec, Inc, Dublin, California; Optical Express, Glasgow, United Kingdom.

Published: August 2020

Purpose: To present outcomes of the United States Food and Drug Administration premarket approval clinical trial of small-incision lenticule extraction (SMILE) for the correction of myopia and astigmatism.

Design: Prospective, multicenter clinical trial.

Participants: The study included 357 eyes of 357 patients treated with SMILE (50 for myopia and 307 for myopia with astigmatism). Preoperative sphere ranged between -1.00 and -10.00 diopters (D), with manifest spherical equivalent (MSE) of up to -11.50 D and refractive cylinder of up to -3.00 D.

Methods: -Participants were followed up for 12 months. Corrected distance visual acuity (CDVA) and uncorrected distance visual acuity (UDVA), stability of the manifest refraction, and vector analysis of refractive cylinder are presented for the 307 eyes treated for myopia with astigmatism. Adverse events (AEs) are presented for all 357 eyes.

Main Outcome Measures: Corrected distance visual acuity, uncorrected distance visual acuity, manifest refraction, astigmatic outcomes, and adverse events.

Results: Of the 307 astigmatic eyes enrolled in the study, 304 were treated successfully. In 3 eyes, the procedure was aborted because of intraoperative suction loss. The mean MSE reduced from -5.39±2.30 D at baseline to -0.01±0.24 D at 12 months. Of all eyes, 95.3% were within 0.50 D of emmetropia at 12 months. The percentage of eyes with UDVA of 20/20 or better was 89.0%. No loss of 2 or more lines of CDVA was observed at the 12-month visit. The refractive cylinder reduced from -1.53±0.70 D at baseline to -0.18±0.31 D at 12 months. The mean correction ratio of refractive cylinder was 0.96±0.16 and a slight undercorrection was apparent for higher attempted corrections of astigmatism. Three intraoperative AEs associated with difficult lenticule removal and resultant cap tear occurred, and all resolved without sequelae at postoperative day 1. During the postoperative period, 8 AEs were recorded, but none of them had significant consequences.

Conclusions: Small-incision lenticule extraction for the treatment of myopia and astigmatism was safe and effective, and the reported AEs had no significant impact on visual acuity. Slight undercorrection of refractive cylinder requires further attention.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ophtha.2020.01.010DOI Listing

Publication Analysis

Top Keywords

refractive cylinder
20
visual acuity
20
myopia astigmatism
16
distance visual
16
small-incision lenticule
12
lenticule extraction
12
extraction smile
8
smile correction
8
correction myopia
8
outcomes united
8

Similar Publications

Purpose: To evaluate the clinical outcome of laser-assisted surgical correction of high hyperopic or mixed astigmatism using small incision intrastromal lenticule rotation (SMILERO) alone or combined with photorefractive keratectomy (PRK).

Methods: This retrospective case series enrolled 25 eyes with high astigmatism that underwent SMILERO surgery. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), uncorrected near visual acuity (UNVA), manifest refraction, central corneal thickness, and corneal higher order aberrations were analyzed before surgery and after 3, 6, and 12 months of follow-up.

View Article and Find Full Text PDF

Purpose: To investigate long-term corneal biomechanical changes in thin corneas after small incision lenticule extraction (SMILE).

Methods: Patients with indications for SMILE were enrolled in this study between November 2017 and March 2018. Patients were matched for age, spherical diopter, cylinder, spherical equivalent (SE), and lenticule thickness (LT), and then categorized into the thin cornea group (preoperative thinnest central corneal thickness [CCT] of 500 µm or less, 32 eyes) or normal cornea group (CCT of greater than 500 µm, 32 eyes).

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcomes of a new FDA approved toric aspheric hydrophobic acrylic intraocular lens.

Patients And Methods: This is a single surgeon, single-arm, on-label, non-randomized, prospective observational study. Thirty patients underwent bilateral cataract surgery (60 eyes) with placement of a Clareon™ Toric IOL in each eye.

View Article and Find Full Text PDF

Visual Performance Correlation with Corneal Aberrometric Profile and Pupil Size After Implantation of a Trifocal Hydrophobic IOL.

Clin Ophthalmol

December 2024

Département de Recherche Clinique, Institut Ophtalmologique de l'Ouest (IOO) Jules Verne, Nantes, France.

Purpose: To evaluate clinical outcomes obtained after cataract surgery involving the implantation of a trifocal hydrophobic intraocular lens (IOL) and to determine if pupil size and the corneal aberrometric profile correlate to visual acuity at different distances.

Methods: 49 patients (98 eyes) underwent bilateral cataract surgery with the placement of FineVision HP IOLs for presbyopia and were assessed at 1- and 3- to 6-months post-surgery. Postoperatively, refraction, monocular and binocular uncorrected and corrected distance visual acuity (UDVA and CDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA) and the binocular defocus curve were measured.

View Article and Find Full Text PDF

Inducing cylindrical and spherical defocus after implantation with new generation intraocular lenses improves intermediate and near visual acuity.

Sci Rep

December 2024

Imaging, Biomechanics and Mathematical Modelling Solutions Lab, Narayana Nethralaya Foundation, Bangalore, India.

An induced cylinder and spherical power after implantation with an extended depth of focus (EDOF) and enhanced monofocal intraocular lens (IOL) could improve distance, intermediate (60 cm) and near (40 cm) visual acuity (VA). In this prospective study, forty eyes with Eyhance EDOF IOL (Johnson and Johnson, USA) and 40 eyes with Vivity EDOF IOL (Alcon Laboratories Inc. USA) were included.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!