Objective: This study aimed to evaluate the influence of the instrument regarding the apical fit and type of the alloy and coronal preflaring procedures in the accuracy of Propex Pixi.

Methods: A total of 40 extracted human single-rooted permanent teeth with apical diameters of 200 µm were selected. A #10 K-file was inserted in the root canal until its end could be observed by a dental microscope to obtain the actual working length (WL). Electronic measurements were performed using Propex Pixi to the root apex ("0.0"). Different file alloys (stainless steel [SS] and nickel titanium [NiTi]) and sizes (#10, #15, and #20) were used before and after coronal flaring. Statistical analysis was performed by a factorial analysis of variance (P≤0.05).

Results: Results showed that the measurements of electronic length (EL) were closer to the actual working length (WL) after coronal flaring (P<0.05). A significant intraclass correlation was observed between EL and WL. In addition, results showed no significant differences between files with different sizes or alloys.

Conclusion: Under the conditions of this study, Propex Pixi demonstrated adequate precision. Its accuracy was enhanced by coronal preflaring procedures regardless of the instrument type used (SS or NiTi) and the apical fit.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7006547PMC
http://dx.doi.org/10.14744/eej.2019.52724DOI Listing

Publication Analysis

Top Keywords

propex pixi
8
coronal preflaring
8
preflaring procedures
8
actual working
8
working length
8
coronal flaring
8
precision propex
4
pixi instruments
4
coronal
4
instruments coronal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!