Transcatheter tricuspid valve-in-ring implantation has emerged as a potential alternative to surgery for high-risk patients with symptomatic severe tricuspid regurgitation that recurs after surgical ring repair. The worldwide experience remains limited. We report a case of unsuccessful transfemoral tricuspid valve-in-ring implantation (using an Edwards SAPIEN 3 valve, Edwards Lifesciences, Irvine, CA) and literature review. The rigidity, open shape, and open configuration of the ring may lead to imperfect positioning, resulting in severe paravalvular leak. Particular attention should be paid to sizing and wire position with respect to the ring while implanting the valve.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7063646PMC
http://dx.doi.org/10.1016/j.cjco.2019.09.005DOI Listing

Publication Analysis

Top Keywords

tricuspid valve-in-ring
12
valve-in-ring implantation
12
unsuccessful transfemoral
8
transfemoral tricuspid
8
literature review
8
tricuspid
4
implantation case
4
case report
4
report literature
4
review transcatheter
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!