This report presents an unusual case of instrumentation failure after posterior fixation of a C2 fracture and reviews currently available treatment alternatives. The patient, a 53-year-old female, initially presented to the emergency department at an outside facility with acute alcohol intoxication and acute neck pain following a fall from a ladder. CT demonstrated bilateral C2 pars fractures and unstable posteroinferior displacement of the posterior elements. She underwent an emergent C2 open-reduction internal fixation (ORIF) at the outside facility with 3.5 mm polyaxial synapse pedicle screws (DePuy Synthes, Switzerland). There were no known complications and the patient was discharged. Two years after the index operation, cervical CT scan at a different facility revealed that although the fracture was fully healed, bilateral tulip caps had detached from the pedicle screw heads at C2. All implants were removed without postoperative complications. Industry review of alternate lag screws approved for the cervical spine demonstrated that there is not currently an ideal implant for fixation of C2 fractures without fusion. Cannulated trauma screws, which are low profile and would have avoided the instrumentation failure seen here, are not currently FDA approved for the cervical spine.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7060438PMC
http://dx.doi.org/10.1155/2020/5824383DOI Listing

Publication Analysis

Top Keywords

instrumentation failure
8
approved cervical
8
cervical spine
8
tulip-screw head
4
head disjunction
4
disjunction posterior
4
posterior fracture
4
fixation
4
fracture fixation
4
fixation instrumentation
4

Similar Publications

A basicervical femoral fracture is a relatively uncommon type of proximal femoral fracture. However, as the proportion of proximal femoral fractures rises in conjunction with the aging of society, the absolute number of patients with basicervical femoral fractures is also increasing. Nevertheless, the optimal surgical methods for the treatment of basicervical femoral fractures remain a topic of debate.

View Article and Find Full Text PDF

The study aimed to evaluate a newly designed semicircular implant for the fixation of Vancouver Type B1 periprosthetic femoral fractures (PFFs) in total hip arthroplasty (THA) patients. To determine its strength and clinical applicability, the new implant was compared biomechanically with conventional fixation methods, such as lateral locking plate fixation and a plate combined with cerclage wires. : Fifteen synthetic femur models were used in this biomechanical study.

View Article and Find Full Text PDF

Valgus-impacted femoral neck fractures (OTA 31B1.1 and 31B1.2) are considered stable fractures with favorable outcomes compared to displaced fractures.

View Article and Find Full Text PDF

Objective: Peritoneal dialysis(PD)-associated peritonitis is a common and major complication of PD and the most common cause of technical failure of PD. The presence of bacterial biofilm may be an important factor leading to refractory or recurrence of peritonitis. To investigate the formation and characteristics of bacterial biofilms on PD catheters after peritonitis-associated catheter removal.

View Article and Find Full Text PDF

Objective: To compare the cyclic fatigue resistance of nickel-titanium files made by 3 new heat treatment in simulated S-shaped root canals at different temperatures.

Methods: Gold heat-treated nickel-titanium files TruNatomy (25 mm, tip size 26#/0.04) and ProTaper Gold (25 mm, tip size 25#/0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!