Objective: The neck is a very attractive measurement location for multimodal physiological monitoring, since it offers the possibility of extracting clinically relevant parameters, which cannot be obtained from other body locations, such as lung volumes. It is for this reason that obtaining PPG from the neck would be of interest. PPG signals, however, are very susceptible to artifacts which greatly compromise their quality. But the extent of this is going to depend on, the nature of the artifacts and the strength of the sensed signal, both of which are location dependent. This paper presents for the first time the characterization of artifacts affecting neck PPG signals.

Methods: Neck PPG data was recorded from 19 participants, who performed ten different activities to deliberately introduce common artifacts. 41 PPG features were extracted and statistically analyzed to investigate which ones showed the greatest ability to differentiate normal PPG from each artifact. A customized minimum Redundancy Maximum Relevance (mRMR) feature selection approach was implemented, to select the top 10 features.

Results: Artifacts caused by Swallowing, Yawning and Coughing exhibited larger Spectral Entropy, Average Power and smaller Spectral Kurtosis, than normal PPG. Head movement artifacts, also demonstrated highly disordered and noisy frequency spectra, and were characterized by having larger and irregular time domain features. In addition, the analysis showed that different respiratory states that could be of clinical interests, such as presence of apneas, were also distinguishable from sources of interference.

Significance: These findings are important for the development of PPG denoising algorithms and subsequent obtention of biomarkers of interest, or alternatively for applications where the events of interest are the artifacts themselves.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2020.2972378DOI Listing

Publication Analysis

Top Keywords

ppg
8
neck ppg
8
normal ppg
8
artifacts
7
neck
5
characterization artifact
4
artifact signals
4
signals neck
4
neck photoplethysmography
4
photoplethysmography objective
4

Similar Publications

Background: The importance of peri-implant soft tissues in maintaining tissue health and aesthetics has been recognized. A thickness of at least 2 mm is considered a protective factor against peri-implantitis. This study assessed clinical outcomes and complications at implant sites following soft tissue augmentation with either palatal free gingival graft (FGG) or palatal pedicle graft (PPG).

View Article and Find Full Text PDF

Accurate and continuous blood glucose monitoring is essential for effective diabetes management, yet traditional finger pricking methods are often inconvenient and painful. To address this issue, photoplethysmography (PPG) presents a promising non-invasive alternative for estimating blood glucose levels. In this study, we propose an innovative 1-second signal segmentation method and evaluate the performance of three advanced deep learning models using a novel dataset to estimate blood glucose levels from PPG signals.

View Article and Find Full Text PDF

Aims: Smartphone applications for heart rate and rhythm assessment are increasingly used for the management of atrial fibrillation (AF). Although the use of a photoplethysmography (PPG)-based smartphone application with subsequent (tele)consultations for AF management has been proven feasible in the TeleCheck-AF project, specific needs, and expectations of patients with AF are unclear. The aim of this study is to evaluate patients' perspectives on the use of remote PPG-based electronical health (eHealth) integrated in regular care pathways for AF.

View Article and Find Full Text PDF

Pyoderma gangrenosum (PG) is the second most common skin manifestation reported in patients with inflammatory bowel disease (IBD). We performed a single-institution, retrospective study to summarize the clinical features and examine effective treatment regimens and outcomes of PG in IBD patients. We identified 45 patients who presented to our institute between January 1, 2002 and December 31, 2021 with the following criteria: (1) diagnosed with an active PG (ICD9: 686.

View Article and Find Full Text PDF

X-ray crystallography is one of the leading tools to analyze the 3-D structure, and therefore, function of proteins and other biological macromolecules. Traditional methods of mounting individual crystals for X-ray diffraction analysis can be tedious and result in damage to fragile protein crystals. Furthermore, the advent of multi-crystal and serial crystallography methods explicitly require the mounting of larger numbers of crystals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!