Background: This study aimed to explore the performance of Revolution CT virtual monoenergetic images (VMI) combined with the multi-material artifact reduction (MMAR) technique in reducing metal artifacts in oral and maxillofacial imaging.

Results: There were significant differences in image quality scores between VMI + MMAR images and VMI+MARS (multiple artifact reduction system) images at each monochromatic energy level (p = 0.000). Compared with the MARS technology, the MMAR technology further reduced metal artifacts and improved the image quality. At VMI and VMI, the SD, CNR, and AI in the Revolution CT group were significantly lower than in the Discovery CT, but no significant differences in these parameters were found between two groups at VMI, VMI, and VMI (p > 0.05). The attenuation was comparable between two groups at any energy level (p > 0.05).

Conclusions: Compared with the MARS reconstruction technique of Discovery CT, the MMAR technique of Revolution CT is better to reduce the artifacts of dental implants in oral and maxillofacial imaging, which improves the image quality and the diagnostic value of surrounding soft tissues.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7058730PMC
http://dx.doi.org/10.1186/s13244-020-0836-1DOI Listing

Publication Analysis

Top Keywords

artifact reduction
12
mmar technique
12
image quality
12
vmi vmi
12
multi-material artifact
8
reduction mmar
8
technique revolution
8
metal artifacts
8
oral maxillofacial
8
energy level
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!