Purpose: The objective was to characterize both systematic and random errors in Positron Emission Tomography (PET)-based Y three-dimensional (3D) dose quantification.

Methods: A modified NEMA-IEC phantom was used to emulate Y-microsphere PET imaging conditions: sphere activity concentrations of 1.6 and 4.8 MBq/cc, sphere-to-background ratios of 4 and 13, and sphere diameters of 13, 17, and 37 mm. PET data were acquired using a GE D690 PET/CT scanner for 300 min on days 0-11. The data were downsampled to 60-5 min for multiple realizations to evaluate the count starvation effect. The image reconstruction algorithm was 3D-OSEM with PSF + TOF modeling; the parameters were optimized for dose-volume histogram (DVH), as a Y 3D dose quantification. Y-PET images were converted to dose maps using the local deposition method, then the sphere DVHs were calculated. The ground truth for the DVH was calculated using convolution method. Dose linearity was evaluated in decaying Y activity (reduced count rate and total count) and decreasing acquisition durations (reduced total count only). Finally, the impacts of the low 32-ppm positron yield on PET-based 3D Y-dose quantification were evaluated; the bias and variability of resulting DVHs were characterized.

Results: We observed nonlinear errors that depended on the Y activity (count rate) and not on the total true prompt counts. These nonlinear errors in mean dose underestimated the measured mean dose by> 20% for a measured dose range of 40-230 Gy; although the shapes of the DVH were not altered. Compensation based on empirical models reduced the nonlinearity errors to be within 5% for measured dose range of 40-230 Gy. In contrast, the errors due to nonuniformity introduced by image noise dominated the systematic errors in the DVH and stretched the DVH on both tails. For the 37-mm sphere, the magnitude of errors in D increased from -25% to -36% when acquisition duration was decreased from 300 to 10 min. The effect of image noise on DVH was more extensive in smaller spheres; for the 17-mm sphere, the magnitude of errors in D increased from -29% to -45% acquisition duration was decreased from 300 to 10 min. For the 37-mm sphere, the errors in D increased from +3.5% to only +10.5% when the acquisition duration was decreased from 300 to 10 min; in the 17-mm sphere, the errors in D were 6.5% for both 300- and 10-min sphere images.

Conclusions: Count-starved Y-PET data introduce both systematic and random errors. The systematic error increases the apparent nonuniformity of the DVH, while the random error increases the uncertainty in the DVH. The systematic errors were larger than the random errors. Lower count rate of Y-PET also introduces systematic bias, which is scanner specific. The errors of bias-compensated mean tumor dose were <10% when Y-PET scan time was >15 min/bed for tumors >37 mm. D and D were the most stable dose metrics. An acquisition duration of 30 min is recommended to keep the random errors < 10% for a typical tumor with sphere equivalent diameter >17 mm and average tumor dose >40 Gy.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14117DOI Listing

Publication Analysis

Top Keywords

random errors
16
acquisition duration
16
errors
15
systematic random
12
dose
12
count rate
12
measured dose
12
errors increased
12
duration decreased
12
decreased 300
12

Similar Publications

Background: Delirium is a common and serious syndrome of acute brain dysfunction associated with negative outcomes. Melatonin may have a role in delirium prevention for critically ill adults based on data from non-critically ill patient populations. Our objective was to assess the feasibility of a multi-centre, randomized, placebo-controlled trial testing the hypothesis that low-dose melatonin prevents delirium in critically ill adults.

View Article and Find Full Text PDF

A ten-year prospective randomized trial comparing non-operative treatment to hook plate fixation for Rockwood III acromio-clavicular dislocation.

Orthop Traumatol Surg Res

January 2025

Faculté de médecine Université Laval, Pavillon Ferdinand-Vandry, 1050 avenue de la Médecine, Québec, Québec G1V 0A6, Canada; Département de chirurgie Université Laval, Pavillon Ferdinand-Vandry, 1050 avenue de la Médecine, Québec, Québec G1V 0A6, Canada; Centre de Recherche CHU de Québec Université Laval, Axe Médecine Régénératrice, 2705 Bd Laurier, Québec, QC G1V 4G2, Canada. Electronic address:

Background: Despite new modern surgical techniques, no short-term functional differences have been demonstrated between operative and non-operative treatment for acute acromio-clavicular (AC) dislocations Rockwood III-V. Few studies describe the long-term clinical and radiological results of these lesions. We aim to compare the one- and ten-year outcome of patients either treated non-operatively or with hook plate fixation for Rockwood III AC lesions.

View Article and Find Full Text PDF

The comprehensive identification of peaks in untargeted lipidomics using LC-MS/MS remains a significant challenge. Confidence in lipid annotation can be greatly improved by integrating a highly accurate machine learning-based retention time prediction model. Such an approach enables the identification of lipids for understanding pathogenic mechanisms, biomarker discovery, and drug screening.

View Article and Find Full Text PDF

Modeling Optical Coherence Tomography (OCT) images is crucial for numerous image processing applications and aids ophthalmologists in the early detection of macular abnormalities. Sparse representation-based models, particularly dictionary learning (DL), play a pivotal role in image modeling. Traditional DL methods often transform higher-order tensors into vectors and then aggregate them into a matrix, which overlooks the inherent multi-dimensional structure of the data.

View Article and Find Full Text PDF

Taking the titanium alloy wing-body connection joint at the rear beam of a certain type of aircraft as the research object, this study analyzed the failure mechanism and verified the structural safety of the wing-body connection joint under actual flight loads. Firstly, this study verified the validity of the loading system and the measuring system in the test system through the pre-test, and the repeatability of the test was analyzed for error to ensure the accuracy of the experimental data. Then, the test piece was subjected to 400,000 random load tests of flight takeoffs and landings, 100,000 Class A load tests, and ground-air-ground load tests, and the test piece fractured under the ground-air-ground load tests.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!