Significance: The diagnostic depth of photodynamic diagnosis (PDD) for gastric cancer with protoporphyrin IX (PpIX) is limited, which leads to missing intramucosal cancers in screening and surgery.

Aim: The reason is that the excitation light, whose wavelength is determined by the highest absorption peak of PpIX (∼405  nm), is strongly attenuated by mucosal tissues. We investigated an excitation wavelength that can extend the diagnostic depth of PpIX fluorescence at the mucosal subsurface.

Approach: By calculating the depth-dependent intensity of the excitation light in porcine gastric mucosa for each wavelength, relationships among the wavelength, fluorophore depth, and fluorescence intensity were assessed and fluorescence images of PpIX pellets located at different fluorophore depths were compared experimentally by changing the excitation wavelength.

Results: The numerical calculation showed that a 505-nm excitation light provided the highest fluorescence intensities at a fluorophore depth deeper than 1.1 mm. In the fluorescence observation, the fluorescence intensities at fluorophore depths of 0 and 1.0 mm at 405 nm were 5.4  ×  103 and 1.0  ×  103  arb. units, whereas those at 505 nm were 5.3  ×  101 and 1.9  ×  102  arb. units, respectively.

Conclusion: The experimental results suggest that the diagnosis depth of PDD with PpIX for intramucosal cancer can be extended by 505-nm excitation light.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7053433PMC
http://dx.doi.org/10.1117/1.JBO.25.6.063809DOI Listing

Publication Analysis

Top Keywords

excitation light
16
intramucosal cancer
8
photodynamic diagnosis
8
diagnostic depth
8
fluorophore depth
8
fluorophore depths
8
505-nm excitation
8
fluorescence intensities
8
intensities fluorophore
8
fluorescence
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!