Aim: Due to reduced PaO , aircrafts at cruising altitudes are pressurised to a cabin altitude of 2438 m, equivalent to breathing FiO 0.15. Portable oxygen concentrators (POCs) are approved for onboard oxygen supply with lack of evidence, especially in infants. We assessed POCs (continuous-flow cPOC vs. pulsed-flow pPOC) under simulated altitude conditions performing Hypoxic Challenge Testing (HCT).

Methods: In a randomised controlled crossover trial, we included patients <1 year born prematurely. In incidents of hypoxia (SpO  ≤ 85%), oxygen was administered through POC. In patients with a positive hypoxia reversal, HCT was repeated 24 hours later. If hypoxia occurred during the second testing, oxygen was given using the alternative POC.

Results: We randomised 26 patients; 22 patients received allocated intervention (4 dropped out). Mean gestational age 30.4 weeks, mean corrected age 38.2 weeks. Both POCs achieved immediate hypoxia reversal in all cases (SpO cPOC/pPOC 98%/99.4% (95%CI -2.91, 0.01)) without any adverse events. No significant difference was observed in patients with BPD.

Conclusion: Both POCs generated sufficient oxygen to reverse HCT induced hypoxia. Although pPOCs are not recommended in paediatric age, our data suggest their effectiveness even in neonates without any associated adverse events. Future research on pPOCs safety is required to establish recommendations for their use.

Download full-text PDF

Source
http://dx.doi.org/10.1111/apa.15242DOI Listing

Publication Analysis

Top Keywords

portable oxygen
8
oxygen concentrators
8
hypoxic challenge
8
challenge testing
8
randomised controlled
8
controlled crossover
8
crossover trial
8
assessment portable
4
concentrators infants
4
infants undergoing
4

Similar Publications

Kirkendall Effect-Mediated Transformation of ZIF-67 to NiCo-LDH Nanocages as Oxidase Mimics for Multicolor Point-of-Care Testing of β-Galactosidase Activity and .

Anal Chem

January 2025

Yunnan Key Laboratory of Modern Separation Analysis and Substance Transformation, College of Chemistry and Chemical Engineering, Yunnan Normal University, Kunming 650500, Yunnan Province, P. R. China.

Early and portable detection of pathogenic bacteria is crucial for ensuring food safety, monitoring product quality, and tracing the sources of bacterial infections. Moving beyond traditional plate-culture counting methods, the analysis of active bacterial components offers a rapid means of quantifying bacteria. Here, metal-organic framework (MOF)-derived NiCo-layered double hydroxide nanosheets (LDHs), synthesized via the Kirkendall effect, were employed as highly effective oxidase mimics to generate reactive oxygen species (ROS).

View Article and Find Full Text PDF

Background: The circadian rhythm of myocardial infarction (MI) in patients with obstructive sleep apnea (OSA) remains disputable and no studies have directly evaluated the relationship between nocturnal hypoxemia and the circadian rhythm of MI. The aim of the current study was to evaluate the association of OSA and nocturnal hypoxemia with MI onset during the night.

Methods: Patients with MI in the OSA-acute coronary syndrome (ACS) project (NCT03362385) were recruited.

View Article and Find Full Text PDF

Introduction: Marathon running has become increasingly popular among amateur athletes, many of whom maintain speeds of 8-9 km/h. However, existing methods for estimating oxygen consumption (VO) during running and walking-such as the American College of Sports Medicine (ACSM) equations and commercial activity monitors-often lack accuracy and transparency. This study introduces the Hata-Yanagiya Physical Activity Calculation (HYPAC) system, a novel approach for estimating VO using Global Positioning System (GPS) and map data.

View Article and Find Full Text PDF

Fatigue-induced incidents in transportation, aerospace, military, and other areas have been on the rise, posing a threat to human life and safety. The determination of fatigue states holds significant importance, especially through reliable and conveniently available physiological indicators. Here, a portable custom-built fNIRS system was used to monitor the fatigue state caused by nap deprivation.

View Article and Find Full Text PDF

Operando Synchrotron X-Ray Absorption Spectroscopy: A Key Tool for Cathode Material Studies in Next-Generation Batteries.

Adv Sci (Weinh)

January 2025

Institute for Superconducting & Electronic Materials (ISEM), Faculty of Engineering and Information Sciences (EIS), University of Wollongong, Wollongong, NSW, 2500, Australia.

Rechargeable batteries are central to modern energy storage systems, from portable electronics to electric vehicles. The cathode material, a critical component, largely dictates a battery's energy density, capacity, and overall performance. This review focuses on the application of operando X-ray absorption spectroscopy (XAS) to study cathode materials in Li-ion, Na-ion, Li-S, and Na-S batteries.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!