Objectives: To compare the stability of NT2B clavicle fractures fixed with either a hook plating (HP), Superior Plating with Suture Augmentation (SPSA), or dual orthogonal plating (DP) with the hypothesis that DP would provide increased multiplanar stability across NT2B fractures.

Methods: NT2B distal clavicle fractures were created in cadaveric specimens and fixed using (1) HP, (2) SPSA, or (3) DP. Specimens were cyclically loaded in 3 different planes of motion: (1) anteroposterior (AP), (2) superior-inferior, and (3) axial rotation while displacement was continually recorded. Afterward, a superiorly directed load was applied to the clavicle. Load to failure, stiffness, and mode of failure were recorded.

Results: During AP loading, clavicles fixed with a DP had significantly lower mean posterior displacement compared to those fixed with SPSA at every 100-cycle interval of testing, P < 0.01. During inferior-superior loading, specimens fixed with a DP had less superior displacement than specimens fixed with an HP and SPSA, reaching significance at the 500-700 cycles of testing. There was no significant difference in axial rotation stability or load to failure between the 3 fixation techniques.

Conclusions: Orthogonally placed minifragment plates provide improved stability against anterior displacement with no significant difference in superior stability, axial rotational stability, stiffness, or load to failure. Further clinical studies are needed to confirm the long-term stability of dual plating and determine the risks and benefits of this novel method of distal clavicle fixation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BOT.0000000000001752DOI Listing

Publication Analysis

Top Keywords

distal clavicle
12
clavicle fractures
12
specimens fixed
12
fixed spsa
12
load failure
12
stability nt2b
8
axial rotation
8
stability
7
fixed
6
clavicle
5

Similar Publications

Background: Clavicle fractures associated with ipsilateral coracoid process fractures are very rare, with limited literature reporting only a few cases. This study reports on 27 patients with ipsilateral concomitant fractures of the clavicle and coracoid process who were followed for more than 12 months.

Material And Methods: This retrospective study reviewed the charts of skeletally mature patients with traumatic ipsilateral clavicle and coracoid process fractures treated at the authors' institution.

View Article and Find Full Text PDF

Objective: To compare the effectiveness of clavicular hook plates and Endobutton plates in treating unstable distal clavicle fractures (UDCFs).

Methods: Data from 95 patients with UDCFs (Neer II and V types) were retrospectively analyzed. Among them, 55 cases were treated with clavicular hook plates (control group), and 40 cases with Endobutton plates (research group).

View Article and Find Full Text PDF

Subacromial Distal Clavicle Dislocation with Associated Scapular Spine Fracture: A Case Report.

J Orthop Case Rep

January 2025

Vanderbilt University Medical Center, Department of Orthopaedic Surgery, 1215 21st Ave S, Suites 3200 and 4200, Nashville, TN, 37232, 2John Hopkins University.

Introduction: Inferior or subacromial dislocation of the distal clavicle is a rare entity. Previous reports of this injury pattern have largely been limited to Rockwood VI acromioclavicular joint (AC) dislocations, with the distal clavicle located in the subcoracoid position. Few case reports have been described with the inferior clavicle being located in the subacromial space, and these have all been previously associated with clavicle fractures.

View Article and Find Full Text PDF

Background: The incidence of revision shoulder arthroplasty continues to rise, and infection is a common indication for revision surgery. Treatment of periprosthetic joint infection (PJI) in the shoulder remains a controversial topic, with the literature reporting varying methodologies, including the use of debridement and implant retention, single-stage and 2-stage surgeries, antibiotic spacers, and resection arthroplasty. Single-stage revision has been shown to have a low rate of recurrent infection, making it more favorable because it precludes the morbidity of a 2-stage operation.

View Article and Find Full Text PDF

Distal augmentation in unstable distal clavicle fractures: a retrospective cohort study of 101 cases.

Arch Orthop Trauma Surg

January 2025

The Department of Orthopaedic Surgery, Kaohsiung Chang Gung Memorial Hospital, Chang Gung University College of Medicine, 123, Ta Pei Road, Niao Sung Dist, Kaohsiung, Taiwan.

Introduction: The optimal management strategy for unstable distal clavicular fractures remains controversial. Recent studies on plate techniques have reported good-to-excellent outcomes with no serious complications. The questions are that: (1) Does the use of wire augmentation with locking plate in distal part (distal wire augmentation) reduce radiographic loss of reduction (RLOR) and get earlier bony union in distal clavicular fractures? (2) Which fixation methods are associated with a higher incidence of acromioclavicular (AC) joints arthritis or subluxation? We collected and analyzed clinical studies on different plate fixation methods for unstable fractures to identify the best surgical modality.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!