AI Article Synopsis

  • The Lancet Commission on Hypertension emphasizes the importance of improving blood pressure measurement accuracy by using validated devices, as many currently available devices lack proper accuracy testing.
  • Weak regulations allow inaccurate devices to be sold without validation, leading to poor hypertension diagnosis and treatment globally.
  • The Commission urges for mandatory independent validation of BP devices and recommends developing standards for new technologies, while supporting the establishment of accessible lists of accurate devices for consumers and healthcare professionals.

Article Abstract

The Lancet Commission on Hypertension identified that a key action to address the worldwide burden of high blood pressure (BP) was to improve the quality of BP measurements by using BP devices that have been validated for accuracy. Currently, there are over 3 000 commercially available BP devices, but many do not have published data on accuracy testing according to established scientific standards. This problem is enabled through weak or absent regulations that allow clearance of devices for commercial use without formal validation. In addition, new BP technologies have emerged (e.g. cuffless sensors) for which there is no scientific consensus regarding BP measurement accuracy standards. Altogether, these issues contribute to the widespread availability of clinic and home BP devices with limited or uncertain accuracy, leading to inappropriate hypertension diagnosis, management and drug treatment on a global scale. The most significant problems relating to the accuracy of BP devices can be resolved by the regulatory requirement for mandatory independent validation of BP devices according to the universally-accepted International Organization for Standardization Standard. This is a primary recommendation for which there is an urgent international need. Other key recommendations are development of validation standards specifically for new BP technologies and online lists of accurate devices that are accessible to consumers and health professionals. Recommendations are aligned with WHO policies on medical devices and universal healthcare. Adherence to recommendations would increase the global availability of accurate BP devices and result in better diagnosis and treatment of hypertension, thus decreasing the worldwide burden from high BP.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7039279PMC
http://dx.doi.org/10.26633/RPSP.2020.21DOI Listing

Publication Analysis

Top Keywords

devices
10
commission hypertension
8
accuracy standards
8
worldwide burden
8
burden high
8
accurate devices
8
accuracy
6
[lancet commission
4
hypertension
4
hypertension group
4

Similar Publications

We demonstrate that performing anatomical pulmonary resection by video-assisted thoracoscopic surgery without staplers or energy devices is feasible. This technique is an alternative for surgeons with limited access to expensive technologies.

View Article and Find Full Text PDF

Introduction: Since the dawn of the new millennium, Candida species have been increasingly implicated as a cause of both healthcare-associated as well as opportunistic yeast infections, due to the widespread use of indwelling medical devices, total parenteral nutrition, systemic corticosteroids, cytotoxic chemotherapy, and broad-spectrum antibiotics. Candida tropicalis is a pathogenic Candida species associated with considerable morbidity, mortality, and drug resistance issues on a global scale.

Methodology: We report a case of a 43-year-old man who was admitted to our hospital for further management of severe coronavirus disease 2019 (COVID-19) pneumonia.

View Article and Find Full Text PDF

Introduction: Hemoperfusion (HP), a blood filtration method targeting the removal of toxins and inflammatory elements, was investigated in this study. The objective was to present the observations in four individuals with confirmed COVID-19 who underwent several rounds of HP utilizing the HA330 cartridge at a hospital in Indonesia.

Case Studies: We report four cases of COVID-19 patients who underwent HP.

View Article and Find Full Text PDF

Background: Point-of-care ultrasound (POCUS) can be used in a variety of clinical settings and is a safe and powerful tool for ultrasound-trained healthcare providers, such as physicians and nurses; however, the effectiveness of ultrasound education for nursing students remains unclear. This prospective cohort study aimed to examine the sustained educational impact of bladder ultrasound simulation among nursing students.

Methods: To determine whether bladder POCUS simulation exercises sustainably improve the clinical proficiency regarding ultrasound examinations among nursing students, evaluations were conducted before and after the exercise and were compared with those after the 1-month follow-up exercise.

View Article and Find Full Text PDF

Background: Management of the extensive soft tissue injuries remains a significant challenge in orthopedic and plastic reconstructive surgery. Since the thumb is responsible for 40% of the functions of the hand, saving and reconstructing a mangled thumb is essential for the patient's future.

Case Presentation: This case report describes the management of a severe occupational thumb injury in a 25-year-old white Persian male who sustained an occupational injury to his left thumb, resulting in extensive burn, crush injury to the distal and proximal phalanx, and severe soft tissue damage to the first metacarpal, thenar, and palmar areas.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!