Purpose: To determine the test-retest reliability and diagnostic accuracy of a binocular optical coherence tomography (OCT) prototype (Envision Diagnostics, El Segundo, California, USA) for pupillometry.

Design: Assessment of diagnostic reliability and accuracy.

Methods: Fifty participants with relative afferent pupillary defects (RAPDs) confirmed using the swinging flashlight method (mean age 49.6 years) and 50 healthy control subjects (mean age 31.3 years) were examined. Participants twice underwent an automated pupillometry examination using a binocular OCT system that presents a stimulus and simultaneously captures OCT images of the iris-pupil plane of both eyes. Participants underwent a single examination on the RAPDx (Konan Medical, Irvine, California, USA), an automated infrared pupillometer. Pupil parameters including maximum and minimum diameter, and anisocoria were measured. The magnitude of RAPD was calculated using the log of the ratio of the constriction amplitude between the eyes. A pathological RAPD was above ±0.5 log units on both devices.

Results: The intraclass correlation coefficient was >0.90 for OCT-derived maximum pupil diameter, minimum pupil diameter, and anisocoria. The RAPDx had a sensitivity of 82% and a specificity of 94% for detection of RAPD whereas the binocular OCT had a sensitivity of 74% and specificity of 86%. The diagnostic accuracy of the RAPDx and binocular OCT was 88% (95% confidence interval 80%-94%) and 80% (95% confidence interval 71%-87%) respectively.

Conclusions: Binocular OCT-derived pupil parameters had excellent test-retest reliability. The diagnostic accuracy of RAPD was inferior to the RAPDx and is likely related to factors such as eye movement during OCT capture. As OCT becomes ubiquitous, OCT-derived measurements may provide an efficient method of objectively quantifying the pupil responses.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2020.02.013DOI Listing

Publication Analysis

Top Keywords

diagnostic accuracy
12
binocular oct
12
automated pupillometry
8
binocular optical
8
optical coherence
8
coherence tomography
8
test-retest reliability
8
reliability diagnostic
8
california usa
8
participants underwent
8

Similar Publications

Early cancer detection substantially improves the rate of patient survival; however, conventional screening methods are directed at single anatomical sites and focus primarily on a limited number of cancers, such as gastric, colorectal, lung, breast, and cervical cancer. Additionally, several cancers are inadequately screened, hindering early detection of 45.5% cases.

View Article and Find Full Text PDF

Aim: Ultrasonography (US) has shown accuracy in imaging healthy periodontium. This study aims to evaluate the feasibility and accuracy of US for estimating dimensions of inflamed periodontium induced by ligature and bacteria.

Methods: Periodontal tissues of maxillary as well as mandibular premolars and molars in six female mini pigs were treated with ligature and three strains of bacteria for 4-10 weeks.

View Article and Find Full Text PDF

This study investigates the prognostic value of serum biomarkers PD-L1 and IGFBP-2 in patients with esophageal carcinoma. It finds a significant positive correlation between these biomarkers and established tumor markers CEA and CYFRA21-1. The 3-year survival rate for the patient cohort was 45.

View Article and Find Full Text PDF

Purpose: The study aimed to develop a deep learning model for rapid, automated measurement of full-spine X-rays in adolescents with Adolescent Idiopathic Scoliosis (AIS). A significant challenge in this field is the time-consuming nature of manual measurements and the inter-individual variability in these measurements. To address these challenges, we utilized RTMpose deep learning technology to automate the process.

View Article and Find Full Text PDF

Enhancing newborn screening sensitivity and specificity for missed NICCD using selected amino acids and acylcarnitines.

Orphanet J Rare Dis

January 2025

Department of Genetics and Metabolism, Children's Hospital of Zhejiang University School of Medicine, National Clinical Research Center for Child Health, No. 3333 Binsheng Road, Binjiang District, Hangzhou, 310053, Zhejiang, China.

Purpose: To enhance the detection rate of Neonatal Intrahepatic Cholestasis caused by Citrin Deficiency (NICCD) through newborn screening (NBS), we analyzed the metabolic profiles of missed patients and proposed a more reliable method for early diagnosis.

Methods: In this retrospective study, NICCD patients were classified into "Newborn Screening" (64 individuals) and "Missed Screening" (52 individuals) groups. Metabolic profiles were analyzed using the non-derivatized MS/MS Kit, and genetic mutations were identified via next-generation sequencing and confirmed by Sanger sequencing.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!