Introduction: We previously reported an antibody MIF-220 that recognizes a specific structure induced on the surface of thrombin-activated E-domain of one fibrin molecule bound with the D-domains of other fibrinogen/fibrin molecules. Utilizing MIF-220, we produced a test kit for cross-linked fibrin degradation products (XDP), LPIA-GENESIS D-dimer (LG-DD), and evaluated basic performance characteristics for clinical application. We then attempted to apply LG-DD to see its eligibility in clinical plasma samples.
Method: The characteristic performances requested for clinical use were studied including limit of quantitation, within-run imprecision, day-to-day imprecision, antigen excess, interference study, and method comparison with LPIAACE-Ddimer (ACE-DD) available on the market.
Results: The performance characteristics were all satisfactory. Extraordinarily high concentrations of XDP are occasionally obtained by ACE-DD in samples with collection problems, but not by LG-DD, indicating that a certain XDP species present in the former was not measured by LG-DD. Structural studies suggested that the "B-b" set of polymerization sites must be involved as well in the maintenance of cross-linked fibrin in vivo.
Conclusion: LG-DD was able to measure a wide range of XDP, that is, 0.20-35.0 μg FEU/mL that covers the levels of XDP in most of the clinical samples. LG-DD was found to almost avoid false-positive results noticed in samples as mentioned above, and this feature seems to be preferable to established kits for the measurement of XDP.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7318163 | PMC |
http://dx.doi.org/10.1111/ijlh.13169 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!