Background: FES (Functional Electrical Stimulation) neuroprostheses have long been a permanent feature in the rehabilitation and gait support of people who had a stroke or have a Spinal Cord Injury (SCI). Over time the well-known foot switch triggered drop foot neuroprosthesis, was extended to a multichannel full-leg support neuroprosthesis enabling improved support and rehabilitation. However, these neuroprostheses had to be manually tuned and could not adapt to the persons' individual needs. In recent research, a learning controller was added to the drop foot neuroprosthesis, so that the full stimulation pattern during the swing phase could be adapted by measuring the joint angles of previous steps.

Methods: The aim of this research is to begin developing a learning full-leg supporting neuroprosthesis, which controls the antagonistic muscle pairs for knee flexion and extension, as well as for ankle joint dorsi- and plantarflexion during all gait phases. A method was established that allows a continuous assessment of knee and foot joint angles with every step. This method can warp the physiological joint angles of healthy subjects to match the individual pathological gait of the subject and thus allows a direct comparison of the two. A new kind of Iterative Learning Controller (ILC) is proposed which works independent of the step duration of the individual and uses physiological joint angle reference bands.

Results: In a first test with four people with an incomplete SCI, the results showed that the proposed neuroprosthesis was able to generate individually fitted stimulation patterns for three of the participants. The other participant was more severely affected and had to be excluded due to the resulting false triggering of the gait phase detection. For two of the three remaining participants, a slight improvement in the average foot angles could be observed, for one participant slight improvements in the averaged knee angles. These improvements where in the range of 4at the times of peak dorsiflexion, peak plantarflexion, or peak knee flexion.

Conclusions: Direct adaptation to the current gait of the participants could be achieved with the proposed method. The preliminary first test with people with a SCI showed that the neuroprosthesis can generate individual stimulation patterns. The sensitivity to the knee angle reset, timing problems in participants with significant gait fluctuations, and the automatic ILC gain tuning are remaining issues that need be addressed. Subsequently, future studies should compare the improved, long-term rehabilitation effects of the here presented neuroprosthesis, with conventional multichannel FES neuroprostheses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7048130PMC
http://dx.doi.org/10.1186/s12984-020-0640-7DOI Listing

Publication Analysis

Top Keywords

joint angles
12
multichannel fes
8
neuroprosthesis
8
drop foot
8
foot neuroprosthesis
8
learning controller
8
physiological joint
8
test people
8
neuroprosthesis generate
8
stimulation patterns
8

Similar Publications

Elbow rotation affects the accuracy of rotational formulas: validation of a modified method.

BMC Musculoskelet Disord

January 2025

Pediatric Orthopedic Hospital, Honghui Hospital, Xi'an Jiao tong University, Xi'an, 710000, China.

Background: Supracondylar humerus fractures (SCHFs) are the most common elbow fractures in children and are typically treated with closed reduction and Kirschner pin fixation. However, varying degrees of residual rotational displacement may remain after closed reduction. Several methods exist to assess rotational displacement, but none account for the effect of elbow rotation on the results.

View Article and Find Full Text PDF

Low back pain (LBP) is a highly prevalent disorder frequently related to biomechanical alterations. Movement pattern assessments have a role in the rehabilitation management of patients with LBP; however, a precise assessment is challenging in routine clinical settings. Thus, this study aims to assess the biomechanical alterations related to LBP through the development and application of an innovative assessment tool named CameraLab.

View Article and Find Full Text PDF

The aim of this comparative observational study is to evaluate and compare the size and position of the condyle among male and female patients with different skeletal patterns in the anterior-posterior dimension using cone beam computed tomography (CBCT) images. CBCT images of 120 patients, all prepared for other treatment purposes under the same conditions, were included in the study. The patients were classified into three groups-class I, class II, and class III-based on ANB angles and Wits analysis.

View Article and Find Full Text PDF

The acromioclavicular (AC) joint, an essential element in the complex biomechanics of the shoulder, displays a diverse range of anatomical variations among individuals. This review aimed to study and present these variations. A detailed search was conducted on the PubMed medical database by using the terms "acromioclavicular joint variations".

View Article and Find Full Text PDF

Background: The study focused on kinematically aligned total knee arthroplasty (KA TKA). It identified which coronal plane alignment of the knee (CPAK) types are associated with a higher proportion of medial deviation of the 6° prosthetic trochlear groove (PTG) relative to the quadriceps' line of pull and whether medial deviation adversely affected the Forgotten Joint Score (FJS). The research calculated the minimum PTG angle required to prevent medial deviation by at least 2° in all patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!