Unlabelled: PRéCIS:: This eye movement perimetry (EMP) study describes the development of saccadic reaction time (SRT)-based visual field plots, which could effectively display the presence, location, and extent of glaucomatous defects and support clinical decision-making.

Purpose: EMP is capable of discriminating normal from glaucomatous visual field defects on the basis of average delays in SRTs. To classify the presence and extent of age-corrected visual field defects, it is required to create SRT-based probability maps.

Aim: The aim of this study was to create visual field probability plots based on SRTs and to evaluate their clinical applicability by 2 glaucoma specialists.

Materials And Methods: The development phase included 95 controls segregated into 5 age-bins to estimate normative limits of SRT. Next, for the testing phase, a set of 28 healthy subjects and 24 glaucoma patients were recruited who underwent standard automated perimetry (SAP) and EMP visual field testing. Fifty-two SAP and EMP plots were presented to 2 glaucoma specialists to classify them as normal or abnormal and to identify the defect location and pattern as 1 or more of 7 predefined categories.

Results: The glaucoma specialists showed a sensitivity of 100% and a specificity of 93% and 96% for identifying normal versus abnormal visual field. For specialists 1 and 2, 85% and 92%, respectively, of EMP reports were assigned to the same category as SAP. The reports that did not agree with SAP were graded to a higher defect pattern. The intermethod agreement for specialists 1 and 2 was κ 0.92 and 0.96, respectively.

Conclusion: SRT-based visual field probability plots provided a comprehensive summary of an individual's visual field status and showed comparable clinical applicability to that of SAP plots.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0000000000001477DOI Listing

Publication Analysis

Top Keywords

visual field
36
visual
9
field plots
8
standard automated
8
automated perimetry
8
eye movement
8
movement perimetry
8
srt-based visual
8
field
8
field defects
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!