  • The study examined the surgical outcomes and complications of totally implantable venous port catheters in 3,000 patients, focusing on demographics and clinical features from March 2005 to March 2017.
  • The majority of catheters were inserted via the right internal jugular vein, with a mean duration of stay of 46.7 months and an overall complication rate of 9.6%, mainly infections and thrombosis.
  • The findings suggest that these catheters can improve the quality of life for cancer patients, despite some risks associated with their use.

Article Abstract

Background: This study aims to present our experience with patients who were performed totally implantable venous port catheter.

Methods: Demographics-clinical features, surgical results and complications of 3,000 patients (1,824 males, 1,176 females; mean age 61.04±11.5 years; range, 18 to 88 years) who were performed totally implantable venous port catheter between March 2005 and March 2017 were evaluated retrospectively. Totally implantable venous port catheter indications, complications, catheter duration of stay, reasons for catheter removal and statistical analysis information were reported in detail by reviewing patient records.

Results: Of the cases, the Seldinger method was used in 98%, 1% (n=36) had the catheter inserted with superficial Doppler ultrasonography, while the veins were not found in 1% (n=29) and these patients were excluded from the study. Totally implantable venous port catheter was inserted via right internal jugular vein in 2,095 patients (70%), via right subclavian vein in 470 patients (16%), via left internal jugular vein in 290 patients (10%), and via left subclavian vein in 106 patients (3%). Mean duration of stay for totally implantable venous port catheter was 46.7 months (range, 1 to 78 months). Complications were detected in a total of 288 patients (9.6%), 153 (5.1%) being early and 135 (4.5%) being late. The most common oncologic indication was colorectal cancer. The rate of port removal was 298/3,000 (9.9%) and the main reasons were infection, thrombosis, pain, and end of treatment. Totally implantable venous port catheter was required to be inserted twice in 33 patients (1%) and thrice in 14 patients (0.5%). Totally implantable venous port catheter had malposition in eight patients and all were revised successfully with over the guide method.

Conclusion: Totally implantable venous port catheter may increase the quality of life in cancer patients. Despite possible complications, totally implantable venous port catheter is safe and comfortable for long-term intravenous treatment. Possible complications may be prevented or managed when totally implantable venous port catheter is performed by surgeons with adequate experience.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7018272PMC
http://dx.doi.org/10.5606/tgkdc.dergisi.2018.15299DOI Listing

Publication Analysis

Top Keywords

totally implantable
44
implantable venous
44
venous port
44
port catheter
40
catheter
13
patients
13
port
12
implantable
11
venous
11
totally
10

Similar Publications

Sandblasting and acid etching are common procedures used to treat implant surfaces, enhancing osseointegration and improving clinical success rates. This clinical study aimed to evaluate the long-term outcomes of sandblasted and acid-etched implants. A total of 303 implants were placed in 114 partially and totally edentulous patients using a two-stage surgical technique and an early loading protocol (6-8 weeks).

View Article and Find Full Text PDF

Background: Cochlear implants (CIs) are neuroprosthetic devices which restore hearing in severe-to-profound hearing loss through electrical stimulation of the auditory nerve. Current CIs use an externally worn audio processor. A long-term goal in the field has been to develop a device in which all components are contained within a single implant.

View Article and Find Full Text PDF

Background: Early detection of peri-implant bone defects can improve long-term durability of dental implants. By the advances in cone-beam computed tomography (CBCT) scanners and introduction of new algorithms, it is important to find the most efficient protocol for detection of bone defects. This study aimed to assess the efficacy of metal artifact reduction (MAR) and advanced noise reduction (ANR) algorithms for detection of peri-implant bone defects.

View Article and Find Full Text PDF

Totally implantable subcutaneous venous access ports (TISVAPs) are essential for long-term central venous chemotherapy, delivering medication directly into the central veins of patients. While they play a critical role in reducing patient discomfort, TISVAPs pose a notable risk of post-insertion infections-particularly concerning for oncology patients with compromised immune systems due to aggressive treatment regimens. Our research addresses this issue by developing a predictive nomogram to estimate the risk of TISVAP-associated infections.

View Article and Find Full Text PDF

In this study, a method for determining the optimal location and orientation of an implantable piezoelectric accelerometer on the short process of the incus is presented. The accelerometer is intended to be used as a replacement for an external microphone to enable totally implantable auditory prostheses. The optimal orientation of the sensor and the best attachment point are determined based on two criteria-maximum pressure sensitivity sum and minimum loudness level sum.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!