Study Design: This is a prospective, stratified randomized, multicenter, 4-year follow-up study.

Objective: The authors aimed to evaluate the long-term clinical efficacy and safety of CaO-SiO2-P2O5-B2O3 glass ceramics (BGS-7) spacers in 1-level posterior lumbar interbody fusion (PLIF) at a 4-year follow-up.

Summary Of Background Data: According to 1-year follow-up results, BGS-7 spacer showed similar fusion rates and clinical outcomes compared with titanium cage. A long-term follow-up study beyond 2 years is necessary to investigate the status of intervertebral bone graft volumes. Moreover, longer follow-up is mandatory to also evaluate the safety and efficacy of BGS-7 spacers, because they remain in the intervertebral space for a long time.

Materials And Methods: In this prospective, randomized, multicenter, 4-year follow-up study, we evaluated 62 of the 74 patients who underwent 1-level PLIF. During 1-level PLIF, titanium cages filled with autologous local bone were inserted into the control group patients and BGS-7 spacers were inserted to the experimental group patients. Bone fusion was evaluated by plain radiography and thin-section computed tomography. Visual Analog Scale (VAS), the Oswestry Disability Index (ODI), Short Form-36 Health Survey (SF-36), and evaluation of safety were conducted after 48 months.

Results: Computed tomography scan showed a bone fusion rate of 90.6% in the BGS-7 spacer group and 93.3% in the control group, with no significant differences between groups. The BGS-7 spacer group showed a significantly larger area directly fused to the endplate than the control group (P<0.001). The BGS-7 spacer group showed a significant increase in the fused area compared with the titanium group at 1- and 4-year follow-up. The ODI, SF-36, back pain, and lower limb pain in both groups showed significant improvement after surgery, and no significant differences were observed between the groups. Both groups showed no additional adverse events.

Conclusions: The 4-year follow-up study showed similar fusion rates and clinical outcomes in both the BGS-7 spacer and autologous bone with a titanium cage in 1-level PLIF. However, the BGS-7 spacer implants showed a larger area of fusion with the endplates than that of autologous bone with a titanium cage. Therefore, the results demonstrated that the BGS-7 spacer can be considered as a novel intervertebral spacer to achieve successful spinal fusion without safety concerns for long-term use.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7447185PMC
http://dx.doi.org/10.1097/BSD.0000000000000950DOI Listing

Publication Analysis

Top Keywords

bgs-7 spacers
12
bgs-7 spacer
12
control group
12
long-term follow-up
8
glass ceramics
8
ceramics bgs-7
8
titanium cage
8
1-level posterior
8
posterior lumbar
8
lumbar interbody
8

Similar Publications

The bioactive glass-ceramic spacer (BGS)-7, a biosynthetic intervertebral fusion material introduced in 2014, has not been the subject of comparative clinical studies on anterior cervical discectomy and fusion (ACDF) surgery. This study, for the first time, aims to compare the radiological and clinical outcomes of the renewed BGS-7, released in 2019, with those of an allograft spacer. The comparison includes a finite element analysis of the biomechanical properties of each implant, adding a novel dimension to the research.

View Article and Find Full Text PDF
Article Synopsis
  • This study compared the effectiveness of CaO-SiO-PO-BO bioactive glass-ceramics (BGS-7) versus allograft spacers with iliac bone grafts (IBG) in patients undergoing multilevel anterior cervical discectomy and fusion (ACDF) surgeries for neck issues.
  • Patients who received BGS-7 had a fusion rate of 89.5%, while those with IBG had a fusion rate of 92.2%. Both methods showed solid integration with bone, and no significant differences were noted between the two types of spacers.
  • Post-surgery, the BGS-7 group maintained better lower segmental height compared to the
View Article and Find Full Text PDF

A CaO-SiO-PO-BO bioactive glass-ceramic (BGS-7) spacer provides high mechanical stability, produces a chemical bond to the adjacent endplate, and facilitates fusion after spine surgery. This prospective, randomized, single-blind, non-inferiority trial aimed to evaluate the radiographic outcomes and clinical efficacy of anterior cervical discectomy and fusion (ACDF) using a BGS-7 spacer for treating cervical degenerative disorders. Thirty-six patients underwent ACDF using a BGS-7 spacer (Group N), and 40 patients underwent ACDF using polyetheretherketone (PEEK) cages filled with a mixture of hydroxyapatite (HA) and β-tricalcium phosphate (β-TCP) for the treatment of cervical degenerative disorders.

View Article and Find Full Text PDF

In an anterior cervical discectomy and fusion (ACDF), various types of graft materials including autograft, allograft, and synthetic graft have been used to achieve adequate spinal fusion. Allograft spacer is mainly used in cervical fusion, especially in the anterior approach. The synthetic bone graft material BGS-7(CaO-SiO2-P2O5-B2O3, bioactive Glass-Ceramics) can bind with surrounding bone tissue by forming a hydroxyapatite layer bone bridge, leading to faster graft osseointegration.

View Article and Find Full Text PDF

Purpose: The CaO-SiO2-P2O5-B2O3 glass-ceramic (BGS-7) spacer is a recently developed spacer that shows chemical bonding to bone with high mechanical stability. Further, this spacer achieves similar results to those of titanium cages. However, evidence regarding the advantages of the BGS-7 spacer is weak compared to polyetheretherketone (PEEK) cage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!