Objective: To report our experience with isolated pressure regulating balloon (PRB) replacement for artificial urinary sphincter (AUS) malfunction in the setting of PRB herniation.

Methods: A retrospective review of our large single-surgeon male AUS database was completed. We analyzed men with herniated PRBs palpable in the groin within an otherwise intact system. Patients with evidence of AUS fluid loss were excluded. PRBs were replaced in a submuscular location through a lower abdominal incision. Continence was defined as requiring ≤1 pad per day. Cystoscopic improvement of sphincter coaptation was confirmed intraoperatively.

Results: Of the 725 patients who underwent AUS surgery between 2011 and 2019, we identified 23 (3.2%) with PRB herniation and persistent or bothersome stress urinary incontinence who underwent isolated PRB replacement (median age 72 years, interquartile range 66-80). Four of the 23 patients were excluded from the analysis for subsequent explant unrelated to PRB replacement. At a mean follow-up of 21.7 months (range 2-99 months), 94.7% of patients (18/19) noted significant improvement in their stress urinary incontinence, and 78.9% of patients (15/19) achieved continence. Median time between AUS placement and PRB revision was 13 months (interquartile range 6-34 months).

Conclusion: PRB replacement appears to be a safe and effective salvage therapy for AUS patients with PRB herniation and persistent incontinence without mechanical failure. Intraoperative cystoscopic confirmation of enhanced sphincter coaptation appears to be a reliable predictor of treatment success.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.urology.2020.01.036DOI Listing

Publication Analysis

Top Keywords

prb replacement
16
pressure regulating
8
regulating balloon
8
artificial urinary
8
urinary sphincter
8
prb
8
sphincter coaptation
8
prb herniation
8
herniation persistent
8
stress urinary
8

Similar Publications

Narrative review: evolution in device technology and advances in surgical techniques on AMS 800 device in the last 50 years.

Transl Androl Urol

August 2024

Department of Urology, Princess Alexandra Hospital, University of Queensland, Brisbane, QLD, Australia.

Background And Objective: The current AMS 800 artificial urinary sphincter (AUS) device is designed to simulate the function of the biological urinary sphincter to prevent urinary flow through mucosal coaptation, compression, and pressure transmission. The challenges in designing the AMS 800 device involve not only the mechanical operation of the artificial sphincter device but also producing a device that is effective, safe, and durable for patients in the long term. The following article provides a narrative review regarding the evolution and development of the AMS 800 devices over the years and evaluates the advances in surgical techniques relating to AMS 800 implantation.

View Article and Find Full Text PDF

Previously, chitosan reduces the senescence-related phenotypes in human foreskin fibroblasts through the transforming growth factor beta (TGF-β) pathway, and enhances the proliferation and migration capabilities of these cells are demonstrated. In this study, we examined whether the senescence-delaying effect of chitosan could be applied to primary knee-related fibroblasts, such as human synovial membrane derived cells (SCs) and anterior cruciate ligament fibroblasts (ACLs). These two types of cells were obtained from donors who needed ACL reconstruction or knee replacement.

View Article and Find Full Text PDF

Background: Human bone marrow-derived stem cells (hBMDSCs) are well characterized mediators of tissue repair and regeneration. An increasing body of evidence indicates that these cells exert their therapeutic effects largely through their paracrine actions rather than clonal expansion and differentiation. Here we studied the role of microRNAs (miRNAs) present in extracellular vesicles (EVs) from hBMDSCs in tissue regeneration and cell differentiation targeting endometrial stromal fibroblasts (eSF).

View Article and Find Full Text PDF

Sulfidated zero-valent iron (S-ZVI) is an attractive material of permeable reactive barriers (PRBs) for the remediation of contaminated groundwater. However, S-ZVI is prone to be passivated due to the oxidation of reactive and conductive iron sulfide (FeS) shell and the formation of inactive and non-conductive ferric (hydr)oxides, which serve as electron transfer barriers to hinder the electron flow from Fe° core to contaminants. This study thus proposed a novel approach for in-situ reactivation and reuse of micronsized S-ZVI (S-mZVI) in PRB using sulfate-reducing bacteria (SRB) enriched culture to realize long-lasting remediation of Cr(VI)-contaminated groundwater.

View Article and Find Full Text PDF

Objective: To describe immediate and long-term outcomes of pressure regulating balloon (PRB) placement, exchange, and extraction from the lateral retroperitoneum (LR) in male patients receiving an artificial urinary sphincter (AUS).

Methods: A retrospective chart review was performed on all patients that underwent primary AUS placement between 2006 and 2021. All patients had the PRB placed in the LR during the study period.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!