Introduction: In this cohort study, we evaluated whether the particles in exhaled air (PExA) device can be used in conjunction with mechanical ventilation during surgery. The PExA device consists of an optical particle counter and an impactor that collects particles in exhaled air. Our aim was to establish the feasibility of the PExA device in combination with mechanical ventilation (MV) during surgery and if collected particles could be analysed. Patients with and without nonsmall cell lung cancer (NSCLC) undergoing lung surgery were compared to normal breathing (NB) patients with NSCLC.

Methods: A total of 32 patients were included, 17 patients with NSCLC (MV-NSCLC), nine patients without NSCLC (MV-C) and six patients with NSCLC and not intubated (NB). The PEx samples were analysed for the most common phospholipids in surfactant using liquid-chromatography-mass-spectrometry (LCMS).

Results: MV-NSCLC and MV-C had significantly lower numbers of particles exhaled per minute (particle flow rate; PFR) compared to NB. MV-NSCLC and MV-C also had a siginificantly lower amount of phospholipids in PEx when compared to NB. MV-NSCLC had a significantly lower amount of surfactant A compared to NB.

Conclusion: We have established the feasibility of the PExA device. Particles could be collected and analysed. We observed lower PFR from MV compared to NB. High PFR during MV may be due to more frequent opening and closing of the airways, known to be harmful to the lung. Online use of the PExA device might be used to monitor and personalise settings for mechanical ventilation to lower the risk of lung damage.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7008139PMC
http://dx.doi.org/10.1183/23120541.00198-2019DOI Listing

Publication Analysis

Top Keywords

pexa device
20
particles exhaled
12
mechanical ventilation
12
patients nsclc
12
patients
8
particle flow
8
compared normal
8
normal breathing
8
breathing patients
8
exhaled air
8

Similar Publications

Access to distal airway samples to assess respiratory diseases is not straightforward and requires invasive procedures such as bronchoscopy and bronchoalveolar lavage. The particles in exhaled air (PExA) device provides a non-invasive means of assessing small airways; it captures distal airway particles (PEx) sized around 0.5-7 μm and contains particles of respiratory tract lining fluid (RTLF) that originate during airway closure and opening.

View Article and Find Full Text PDF

Background: Delayed onset muscles soreness (DOMS) is an indication of muscle stress and trauma that develops from excessive musculoskeletal loads. Musculoskeletal loads can be measured with wearable devices, but there is limited research on specific training load metrics that most correlate with DOMS after activity.

Purpose: To determine the predictive capabilities of training load variables on the development of lower extremity DOMS in female collegiate soccer athletes throughout an entire season.

View Article and Find Full Text PDF

Background: SARS-CoV-2 has been shown to predominantly infect the airways and the respiratory tract and too often have an unpredictable and different pathologic pattern compared to other respiratory diseases. Current clinical diagnostical tools in pulmonary medicine expose patients to harmful radiation, are too unspecific or even invasive. Proteomic analysis of exhaled breath particles (EBPs) in contrast, are non-invasive, sample directly from the pathological source and presents as a novel explorative and diagnostical tool.

View Article and Find Full Text PDF

Background: Screening decreases mortality among lung cancer patients but is not widely implemented, thus there is an unmet need for an easily accessible non-invasive method to enable early diagnosis. Particles in exhaled air offer a promising such diagnostic tool. We investigated the validity of a particles in exhaled air device (PExA) to measure the particle flow rate (PFR) and collect exhaled breath particles (EBP) to diagnose primary lung adenocarcinoma (LUAD).

View Article and Find Full Text PDF

Aims: Exhaled breath particles have been explored for diagnosing different lung diseases. We recently showed in an experimental model that different cardiac output results in different particle flow rate (PFR) from the airways. Given the well-known close relationship between impaired cardiac function and respiratory failure, we hypothesized that PFR in exhaled air can be used to detect cardiac failure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!