We compared inductively coupled plasma-mass spectrometry (ICP-MS) test results for the analysis of heavy metals (As, Ba, Cd, Hg, Pb, and Se) in pet foods and routine veterinary diagnostic specimens using intralaboratory and interlaboratory comparisons. Four laboratories, 1 principal laboratory and 3 collaborating laboratories, conducted instrument comparison (limit of detection [LOD], limit of quantification [LOQ], and linear dynamic range [LDR] on 24 data sets), in-house method comparison (accuracy and precision on 120 data sets), and interlaboratory comparison (reproducibility on 528 data sets using Horwitz equation analysis). Matrices tested included 2 types of pet food jerky treats (chicken and sweet potato), bovine blood, and bovine liver and kidney. The instrument comparison study confirmed that ICP-MS provided the sensitivity necessary for the analysis of all heavy metals tested at concentrations below the level of concern for routine diagnostic testing. The "in-house" method comparison samples, spiked at low (0.04 µg/g), medium (0.4 µg/g), and high (8.0 µg/g; note: the high validation level spike for mercury was 2 µg/g) concentration levels, indicated that ICP-MS can meet U.S. FDA acceptance criteria for both accuracy (90-105% recovery) and precision (< 6% coefficient of variation). The interlaboratory comparison studies showed that ICP-MS is a reproducible method for the analysis of heavy metals (HorRat value of 0.5-2.0) except for mercury in one laboratory, which used a different sample preparation method (open block rather than microwave digestion). Overall, our study showed that ICP-MS is a reproducible method for the analysis of heavy metals in spite of minor differences in methodology.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7081506PMC
http://dx.doi.org/10.1177/1040638720903115DOI Listing

Publication Analysis

Top Keywords

analysis heavy
16
heavy metals
16
interlaboratory comparison
12
data sets
12
diagnostic specimens
8
inductively coupled
8
coupled plasma-mass
8
plasma-mass spectrometry
8
instrument comparison
8
method comparison
8

Similar Publications

The widespread application of swine-farming wastewater to soil and water is increasingly contributing to heavy metal contamination, posing significant environmental risks. This study investigated the concentrations of eight heavy metals in swine-farming wastewater following different treatment processes, and assessed their ecological risks in Sichuan Province, China. The findings revealed that zinc, copper and nickel exhibited the highest concentrations, potentially causing heavy or strong contamination levels and leading to heavy or slight ecological risks.

View Article and Find Full Text PDF

Development of detection system for lead ions in mixture solutions using UV-Vis measurements with peptide immobilized microbeads.

Sci Rep

January 2025

Faculty of Frontiers of Innovative Research in Science and Technology (FIRST), Konan University, Chuo-ku, Kobe, 650-0047, Hyogo, Japan.

Environmental pollution caused by heavy metals are problems worldwide. In particular, pollution and poisoning by lead ions (Pb) continue to be common and serious problems. Hence, there is a need for a widely usable method to easily detect Pb from solutions containing organic materials from environmental water such as seas, ponds, etc.

View Article and Find Full Text PDF

The current work introduces the hybrid ensemble framework for the detection and segmentation of colorectal cancer. This framework will incorporate both supervised classification and unsupervised clustering methods to present more understandable and accurate diagnostic results. The method entails several steps with CNN models: ADa-22 and AD-22, transformer networks, and an SVM classifier, all inbuilt.

View Article and Find Full Text PDF

Background: Allowing a birth companion is the basic right of a mother and is identified as an important component of respectful maternity care. The implementation of this intervention has been a challenge in heavy-load public health facilities in India.

Local Problem: Despite the proven benefits of the presence of birth companions on maternal-fetal outcomes, there was no policy of allowing birth companions in our hospital.

View Article and Find Full Text PDF

Advancements in functional adsorbents for sustainable recovery of rare earth elements from wastewater: A comprehensive review of performance, mechanisms, and applications.

Adv Colloid Interface Sci

January 2025

School of Metallurgy and Environment, Central South University, Changsha 410083, Hunan, China; Chinese National Engineering Research Center for Control & Treatment of Heavy Metal Pollution, Institute of Environmental Engineering, Central South University, Changsha 410083, Hunan, China.

Rare earth elements (REEs) are crucial metallic resources that play an essential role in national economies and industrial production. The reclaimation of REEs from wastewater stands as a significant supplementary strategy to bolster the REEs supply. Adsorption techniques are widely recognized as environmentally friendly and sustainable methods for the separation of REEs from wastewater.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!