Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer.

J Cataract Refract Surg

IRCCS-G.B. Bietti Foundation (Savini, Balducci, Schiano-Lomoriello), Rome, Studio Oculistico d'Azeglio (Barboni), Bologna, San Raffaele Hospital (Barboni), Milan, Italy; Stein Eye Institute (Hoffer), University of California, Los Angeles, St. Mary's Eye Center (Hoffer), Santa Monica, California, USA.

Published: January 2020

Purpose: To analyze the results of intraocular lens (IOL) power calculation using measurements by a swept-source optical coherence tomography (SS-OCT) optical biometer.

Setting: IRCCS G.B. Bietti Foundation, Rome, Italy.

Design: Evaluation of a diagnostic test instrument.

Methods: Preoperative measurements by the OA-2000 (Tomey Inc.) were taken in a consecutive series of patients undergoing cataract surgery with one IOL model (AcrySof SN60WF; Alcon Laboratories, Inc.). Measurements were entered into the following formulas: Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Holladay 1, Holladay 2, Holladay 2 with axial length adjustment, Kane, Olsen, Panacea, SRK/T, T2, and VRF. When refraction was measured at 1 month postoperatively, the mean arithmetic prediction error, the median absolute error (MedAE), and the percentage of eyes with a error of ±0.50 D or less were calculated after constant optimization.

Results: We enrolled 150 eyes. All formulas yielded excellent outcomes. The MedAE ranged between 0.200 D and 0.259 D, with a statistically significant difference among formulas (P = .0004). The lowest MedAE values were obtained with the Barrett, EVO, Kane, Olsenstandalone, Radial Basis Function (RBF), and T2 formulas. The percentage of eyes with a prediction error of ±0.50 D or less ranged between 80.00% and 90.67%, with a statistically significant difference (P < .0001). The Barrett, EVO, Holladay 2 with axial length adjustment, Kane, RBF, and T2 achieved the highest percentages (≥88%).

Conclusions: Measurements provided by the SS-OCT optical biometer enable accurate IOL power calculation because all formulas yielded a prediction error within 0.50 D in at least 80% of eyes.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2019.08.044DOI Listing

Publication Analysis

Top Keywords

power calculation
12
prediction error
12
intraocular lens
8
measurements swept-source
8
swept-source optical
8
optical coherence
8
coherence tomography
8
optical biometer
8
iol power
8
ss-oct optical
8

Similar Publications

This in-vitro study assessed the influence of the shade of human teeth on the transmission of near-infrared light. A total of 40 teeth were used. After cleaning the root surface and removing cementum, the teeth were sectioned into slices 3 mm thick, with each comprising a portion of the crown (enamel-dentine (ED)) and of the root (dentine only).

View Article and Find Full Text PDF

Unsupervised segmentation of heart sounds from abrupt changes detection.

Comput Biol Med

January 2025

Univ. Grenoble Alpes, CNRS, CHU Grenoble Alpes, Grenoble INP, TIMC-IMAG, La Tronche, France.

Background And Objective: Heart auscultation enables early diagnosis of cardiovascular diseases. Automated segmentation of cardiograms into fundamental heart states can guide physicians to analyze the patient's condition more effectively. In this work, we propose an unsupervised method of segmentation into heart sounds and silences based on the detection of abrupt changes in the signal.

View Article and Find Full Text PDF

Introduction: In response to blood shortages, providers face pressure to conserve blood. No metrics exist to calculate transfusion utility. We describe characteristics of survivors after high-volume resuscitation and evaluate transfusion utility in low-volume and high-volume resuscitation.

View Article and Find Full Text PDF

Mechanistic insight into the decomposition of sulfone compounds in supercritical water.

J Environ Manage

January 2025

State Key Laboratory of Multiphase Flow in Power Engineering (SKLMF), Xi'an Jiaotong University, NO.28 Xianning West Road, Xi'an, 710049, Shaanxi Province, China.

Supercritical water gasification (SCWG) is famous for the clean utilization of organic wastes without SO emission. Investigating the decomposition mechanism of sulfone compounds, the dominant organic sulfur compounds of organic wastes, in supercritical water (SCW) is conducive to the development of SCWG technology. Herein, the comparative decomposition mechanism of phenyl vinyl sulfone (PVS), diphenyl sulfone (DS), and benzo[b]thiophene 1,1-dioxide (BD) are explored via experiments and density functional theoretical (DFT) calculations.

View Article and Find Full Text PDF

A phylogenetic approach to delimitate species in a probabilistic way.

Syst Biol

January 2025

Division of Ecology & Evolution, Research School of Biology, Australian National University, Canberra ACT 0200 Australia.

Different species concepts and their associated criteria have been used to delimit species boundaries, such as the absence of gene flow for the biological species concept and the presence of morphological distinction for the morphological species concept. The need for different delimitation criteria largely reflects the fact that species are generated under various speciation mechanisms. A key question is how to make species delimitation consistent in a species group, especially when we want to delimit the species boundaries over many newly discovered evolutionary lineages and add these new lineages into a comparative analysis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!