An optimization algorithm for dose reduction with fluence-modulated proton CT.

Med Phys

Department of Medical Physics, Faculty of Physics, Ludwig-Maximilians-Universität München, Am Coulombwall 1, 85748, Garching b. München, Germany.

Published: April 2020

Purpose: Fluence-modulated proton computed tomography (FMpCT) using pencil beam scanning aims at achieving task-specific image noise distributions by modulating the imaging proton fluence spot-by-spot based on an object-specific noise model. In this work, we present a method for fluence field optimization and investigate its performance in dose reduction for various phantoms and image variance targets.

Methods: The proposed method uses Monte Carlo simulations of a proton CT (pCT) prototype scanner to estimate expected variance levels at uniform fluence. Using an iterative approach, we calculate a stack of target variance projections that are required to achieve the prescribed image variance, assuming a reconstruction using filtered backprojection. By fitting a pencil beam model to the ratio of uniform fluence variance and target variance, relative weights for each pencil beam can be calculated. The quality of the resulting fluence modulations is evaluated by scoring imaging doses and comparing them to those at uniform fluence, as well as evaluating conformity of the achieved variance with the prescription. For three different phantoms, we prescribed constant image variance as well as two regions-of-interest (ROI) imaging tasks with inhomogeneous image variance. The shape of the ROIs followed typical beam profiles for proton therapy.

Results: Prescription of constant image variance resulted in a dose reduction of 8.9% for a homogeneous water phantom compared to a uniform fluence scan at equal peak variance level. For a more heterogeneous head phantom, dose reduction increased to 16.0% for the same task. Prescribing two different ROIs resulted in dose reductions between 25.7% and 40.5% outside of the ROI at equal peak variance levels inside the ROI. Imaging doses inside the ROI were increased by 9.2% to 19.2% compared to the uniform fluence scan, but can be neglected assuming that the ROI agrees with the therapeutic dose region. Agreement of resulting variance maps with the prescriptions was satisfactory.

Conclusions: We developed a method for fluence field optimization based on a noise model for a real scanner used in pCT. We demonstrated that it can achieve prescribed image variance targets. A uniform fluence field was shown not to be dose optimal and dose reductions achievable with the proposed method for FMpCT were considerable, opening an interesting perspective for image guidance and adaptive therapy.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14084DOI Listing

Publication Analysis

Top Keywords

image variance
24
uniform fluence
24
dose reduction
16
variance
14
pencil beam
12
fluence field
12
fluence
10
dose
8
fluence-modulated proton
8
image
8

Similar Publications

Background And Aim: In dental clinics, disinfecting alginate impression materials is a critical practice to prevent cross-infection. Recently, zinc oxide nanoparticles (ZnO NPs) have been explored for their potential antimicrobial properties, making them promising additives for dental materials. This study investigates the antimicrobial activity of ZnO NPs incorporated into alginate impression materials and assesses the impact on material flow.

View Article and Find Full Text PDF

Aim To compare the quality of obturation using WVC (warm vertical compaction), CLC (cold lateral compaction), injectable (iFill), and Thermafil (GuttaCore) techniques, along with hydraulic condensation of bioceramic (BC) sealer, and evaluating the percentage of gutta-percha (GP), sealer, and voids in simulated internal resorptive lesions, using an advanced stereomicroscope with ImageJ software (National Institutes of Health, Bethesda, MD, USA). Methods and material In this study, 40 freshly extracted maxillary incisors were collected, and endodontic instrumentation was done to working length using hand K-files. Simulated internal resorption cavities were created in the middle-apical third of the roots after horizontal sectioning and were re-cemented.

View Article and Find Full Text PDF

Causal relationships between immune cells, inflammatory factors, and preeclampsia: A two-step, two-sample mendelian randomization study.

J Reprod Immunol

January 2025

Department of Gynecology and Obstetrics, Tianjin Medical University General Hospital, Tianjin, China; Tianjin Key Laboratory of Female Reproductive Health and Eugenics, Tianjin Medical University General Hospital, Tianjin, China. Electronic address:

Background: Preeclampsia (PE) is a complex hypertensive disorder that occurs during pregnancy, with the immune system playing a key role. Although immune modulation is implicated in PE progression, the roles of specific immune cells and inflammatory mediators remain unclear.

Methods: We conducted a two-sample, two-step Mendelian randomization (MR) analysis, primarily using the inverse-variance weighted method, to investigate the causal effect of immune cell traits on PE.

View Article and Find Full Text PDF

Background: Point-of-care ultrasound (POCUS) can be used in a variety of clinical settings and is a safe and powerful tool for ultrasound-trained healthcare providers, such as physicians and nurses; however, the effectiveness of ultrasound education for nursing students remains unclear. This prospective cohort study aimed to examine the sustained educational impact of bladder ultrasound simulation among nursing students.

Methods: To determine whether bladder POCUS simulation exercises sustainably improve the clinical proficiency regarding ultrasound examinations among nursing students, evaluations were conducted before and after the exercise and were compared with those after the 1-month follow-up exercise.

View Article and Find Full Text PDF

Background: Innovation in crop establishment is crucial for wheat productivity in drought-prone climates. Seedling establishment, the first stage of crop productivity, relies heavily on root and coleoptile system architecture for effective soil water and nutrient acquisition, particularly in regions practicing deep planting. Root phenotyping methods that quickly determine coleoptile lengths are vital for breeding studies.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!