Axial Displacements and Removal Torque Changes of Five Different Implant-Abutment Connections under Static Vertical Loading.

Materials (Basel)

Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul 03080, Korea.

Published: February 2020

The aim of this study was to examine the settling of abutments into implants and the removal torque value under static loading. Five different implant-abutment connections were selected (Ext: external butt joint + two-piece abutment; Int-H2: internal hexagon + two-piece abutment; Int-H1: internal hexagon + one-piece abutment; Int-O2: internal octagon + two-piece abutment; Int-O1: internal octagon + one-piece abutment). Ten implant-abutment assemblies were loaded vertically downward with a 700 N load cell at a displacement rate of 1 mm/min in a universal testing machine. The settling of the abutment was obtained from the change in the total length of the entire implant-abutment unit before and after loading using an electronic digital micrometer. The post-loading removal torque value was compared to the initial torque value with a digital torque gauge. The settling values and removal torque values after 700 N static loading were in the following order, respectively: Ext < Int-H1, Int-H2 < Int-O2 < Int-O1 and Int-O2 < Int-H2 < Ext < Int-H1, Int-O1 (α = 0.05). After 700 N vertical static loading, the removal torque values were statistically different from the initial values, and the post-loading values increased in the Int-O1 group and Int-H1 group (α = 0.05) and decreased in the Ext group, Int-H2 group, and Int-O2 group (α = 0.05). On the basis of the results of this study, it should be taken into consideration that a loss of the preload due to the settling effect can lead to screw loosening during a clinical procedure in the molar region where masticatory force is relatively greater.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7040660PMC
http://dx.doi.org/10.3390/ma13030699DOI Listing

Publication Analysis

Top Keywords

removal torque
20
static loading
12
two-piece abutment
12
implant-abutment connections
8
internal hexagon
8
one-piece abutment
8
internal octagon
8
torque values
8
ext int-h1
8
group 005
8

Similar Publications

Objectives: This study aimed to evaluate the osseointegration properties of titanium bone implants coated with carob-mediated calcium hydroxide nanoparticles biomechanically, radiographically, and histologically on rabbit tibias.

Material And Methods: Forty coated and forty uncoated titanium alloy bone implants were inserted into 20 New Zealand rabbits; each tibia received 2 implants. The rabbits were sacrificed after 4 or 8 weeks, and samples were retrieved for biomechanical evaluation through removal torque test to assess the bond between implant and bone, radiographic evaluation through microcomputed tomography analysis to compare the bone-to-implant contact percentage and bone volume of the peri-implant area, scanning electron microscopic and histologic evaluation through hematoxylin and eosin stain.

View Article and Find Full Text PDF

Background: Anterior column realignment is an attractive minimally invasive treatment for sagittal imbalance. Expandable spacers offer controlled tensioning of the anterior longitudinal ligament (ALL) during release, which could optimize correction and anterior column stability. This study investigated the biomechanical and radiographic effects of single-level anterior-to-psoas lumbar interbody fusion (ATP-LIF) with expandable spacers and sequential ALL release.

View Article and Find Full Text PDF

Exploring the impact of blood and bone debris at the implant-abutment interface on peak removal torque: Bridging clinical observations with laboratory evidence.

J Prosthet Dent

November 2024

Associate Professor, State Key Laboratory of Military Stomatology & National Clinical Research Center for Oral Diseases & Shaanxi Engineering Research Center for Dental Materials and Advanced Manufacture, Department of Implant Dentistry, School of Stomatology, The Air Force Medical University, Xi'an, PR China. Electronic address:

Statement Of Problem: Three months after implant placement, difficulty was encountered in unscrewing the healing abutment within the normal torque range (<35 Ncm), resulting in the removal of the implant. Despite all attempts, the implant could not then be separated from the healing abutment. Retrospective analysis of intraoperative photographs revealed the presence of blood or bone debris within the implant before the healing abutment was placed; this was considered to be the likely cause of this phenomenon.

View Article and Find Full Text PDF

Estrogen deficiency is one of several contributing factors to catabolic changes in bone surrounding dental implants, impairing bone repair in defects requiring bone regeneration. Functionalizing bone substitutes is an alternative approach among various strategies to address this challenge. In this study, the aim was to evaluate the effect of functionalizing deproteinized bovine bone (Bio-Oss, BO) with genistein via sonication on peri-implant bone defects in ovariectomized rats.

View Article and Find Full Text PDF

Background: Acetabular labral tear morphology or orientation may influence hip stability.

Hypothesis: A radial tear of the acetabular labrum would result in greater rotational and translational motion compared with a chondrolabral separation.

Study Design: Controlled laboratory study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!