Intraoperative detection of stuck leaflet after implantation of a mechanical aortic valve: a case report.

JA Clin Rep

Department of Anesthesia, New Heart Watanabe Institute, Hamadayama 3-19-11, Suginami-ku, Tokyo, 168-0065, Japan.

Published: February 2020

Background: Mechanical aortic valve dysfunction is a rare but potentially fatal complication. It is usually caused by thrombosis, vegetation, and pannus overgrowth. However, it very rarely occurs immediately after the cardiopulmonary bypass weaning period.

Case Presentation: We describe a case of stuck leaflet after the implantation of a mechanical aortic valve because of residual suture materials interfering with mechanical aortic valve leaflet closure, which is a very rare cause and has not been reported until now.

Conclusion: The residual suture materials for mechanical aortic valve replacement could cause mechanical valve dysfuction indicated as stuck leafet.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7005232PMC
http://dx.doi.org/10.1186/s40981-020-0313-3DOI Listing