Background: Inaccuracies in treatment setup during radiation therapy for breast cancers may increase risks to surrounding normal tissue toxicities, i.e. organs at risks (OARs), and compromise disease control. This study was planned to evaluate the dosimetric and isocentric variations and determine setup reproducibility and errors using an online electronic portal imaging (EPI) protocol.

Methods: A total of 360 EPIs in 60 patients receiving breast/chest wall irradiation were evaluated. Cumulative dose-volume histograms (DVHs) were analyzed for mean doses to lung (V20) and heart (V30), setup source to surface distance (SSD) and central lung distance (CLD), and shifts in anterior-posterior (AP), superior-inferior (SI), and medial lateral (ML) directions.

Results: Random errors ranged from 2 to 3 mm for the breast/chest wall (medial and lateral) tangential treatments and 2-2.5 mm for the anterior supraclavicular nodal field. Systematic errors ranged from 3 to 5 mm in the AP direction for the tangential fields and from 2.5 to 5 mm in the SI and ML direction for the anterior supraclavicular nodal field. For right-sided patients, V20 was 0.69-3.96 Gy, maximum lung dose was 40.5 Gy, V30 was 1.4-3 Gy, and maximum heart dose was 50.5 Gy. Similarly, for left-sided patients, the CLD (treatment planning system) was 25 mm-30 mm, CLD (EPIs) was 30-40 mm, V20 was 0.9-5.9 Gy, maximum lung dose was 45 Gy, V30 was 2.4-4.1 Gy, and maximum heart dose was 55 Gy.

Conclusion: Online assessment of patient position with matching of EPIs with digitally reconstructed radiographs (DRRs) is a useful method in evaluation of interfraction reproducibility in breast irradiation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6994764PMC
http://dx.doi.org/10.1016/j.mjafi.2018.07.008DOI Listing

Publication Analysis

Top Keywords

dosimetric isocentric
8
isocentric variations
8
treatment planning
8
electronic portal
8
portal imaging
8
breast/chest wall
8
medial lateral
8
errors ranged
8
anterior supraclavicular
8
supraclavicular nodal
8

Similar Publications

Purpose: To evaluate the impact of patient setup errors on the dosimetry and radiobiological models of intensity-modulated radiotherapy (IMRT) for esophageal cancer.

Methods And Materials: This retrospective study with 56 patients in thermoplastic mask (TM) and vacuum bag (VB) groups utilized real setup-error (RSE) data from cone-beam CT scans to generate simulated setup-error (SSE) data following a normal distribution. The SSE data were applied to simulate all treatment fractions per patient by shifting the plan isocenter and recalculating the dose.

View Article and Find Full Text PDF

Purpose: To determine the dosimetric effects of set-up errors on boost coverage, and compares skin toxicity of sequential and simultaneous boost techniques for left-sided breast cancer.

Materials And Methods: This retrospective study included 23 early-stage breast cancer cases. Single isocenter HFWBI-SIB(s-SIB), single isocenter HFWBI-SB(s-SB) and dual isocenter HFWBI-SB(d-SB) were planing.

View Article and Find Full Text PDF

Introduction: SRS for the treatment of limited brain metastases (BM) is widely accepted, but there are still limitations in the management of numerous BM. Frameless single-isocenter multitarget SRS is a novel technique that allows for rapid treatment delivery to multiple BM. We report our preliminary clinical, dosimetric, and patient´s shifts outcomes with this technique.

View Article and Find Full Text PDF

This study evaluates the dosimetric impact of arc simulation angular resolution in VMAT-based Single Isocentre Multiple Target (SIMT) SRS, focusing on their dependence on target size, isocentre distance, number of arcs, and arc type. A phantom study analysed angular resolution (0.5°, 1°, 2°) effects on dosimetric accuracy for PTVs of 0.

View Article and Find Full Text PDF

Background/aim: HyperArc (HA) is an automated planning technique enabling single-isocenter brain stereotactic radiotherapy (SRT); however, dosimetric outcomes may be influenced by the planner's expertise. This study aimed to assess the impact of institutional experience on the plan quality of HA-SRT for both single and multiple brain metastases.

Materials And Methods: Twenty patients who underwent HA-SRT for single metastasis between 2020 and 2021 comprised the earlier group, while those treated between 2022 and 2024 constituted the later group.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!