Development of a circulatory mock loop for biventricular device testing with various heart conditions.

Int J Artif Organs

Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, OH, USA.

Published: September 2020

This study aimed to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing. The mock loop was built with dedicated ports suitable for attaching assist devices in various configurations. This biventricular mock loop uses two pneumatic pumps (Abiomed AB5000, Danvers, MA, USA) driven by a dual-output driver (Thoratec Model 2600, Pleasanton, CA, USA). The drive pressures can be individually modified to simulate a healthy heart and left and/or right heart failure conditions, and variable compliance and fluid volume allow for additional customization. The loop output for a healthy heart was tested at 4.2 L/min with left and right atrial pressures of 1 and 5 mm Hg, respectively; a mean aortic pressure of 93 mm Hg; and pulmonary artery pressure of 17 mm Hg. Under conditions of left heart failure, these values were reduced to 2.1 L/min output, left atrial pressure = 28 mm Hg, right atrial pressure = 3 mm Hg, aortic pressure = 58 mm Hg, and pulmonary artery pressure = 35 mm Hg. Right heart failure resulted in the reverse balance: left atrial pressure = 0 mm Hg, right atrial pressure = 30 mm Hg, aortic pressure = 100 mm Hg, and pulmonary artery pressure = 13 mm Hg with a flow of 3.9 L/min. For biventricular heart failure, flow was decreased to 1.6 L/min, left atrial pressure = 13 mm Hg, right atrial pressure = 13 mm Hg, aortic pressure = 52 mm Hg, and pulmonary artery pressure = 18 mm Hg. This mock loop could become a reliable bench tool to simulate a range of heart failure conditions.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0391398820903316DOI Listing

Publication Analysis

Top Keywords

mock loop
20
heart failure
20
left atrial
16
pulmonary artery
16
circulatory mock
8
device testing
8
heart
8
healthy heart
8
failure conditions
8
atrial pressure = 13 mm hg
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!